FDA Device Recalls

Recalls /

#140519

Product

GE Healthcare, SIGNA 3.0T. Product Usage: Magnetic resonance system for diagnostic imaging. The indications for use for the 3.0T Signa VH/i (Signa 3.0T MR System) Transmit/Receive Body Imaging Coil expands the imaging capability of the 3.0T Signa VH/i MR Imaging System. The Transmit/Receive Body Imaging Coil is intended for imaging of the Neck, Spine, Abdomen/Thorax and the extremities.

FDA product code
LNHSystem, Nuclear Magnetic Resonance Imaging
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K013636
Affected lot / code info
00000262893MR1 520694MR3 00000GE0206027 082427040020 00000243366MR2 082427040040 00000256194MR2 409740MR01 00000261205MR9 083027828659711 00000280097MR7 IE1111MR01 00000189273MR6 A5155133 00000272076MR1 YM0222 00000267372MR1 YM0333 00000000004230 YM0451 00000271322MR0 0850270094 00000227450MR7 333MR3TXXX 00000271777MR5 5606MRS3T 00000273539MR7 570844HCM3T 00000227450MR7 333MR3T 000000HDMR5684 CEN03406 00000GE0209032 0853270052IB1 00000260651MR5 10491MRS01

Why it was recalled

When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.

Root cause (FDA determination)

Software design

Action the firm took

GE Healthcare sent an Urgent Medical Device Correction letter dated August 31, 2015. The letter was addressed to Hospital Administrators/Risk Managers, Radiology Department Managers, & Radiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Recalling firm

Firm
GE Healthcare
Address
3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615

Distribution

Distribution pattern
Worldwide Distribution - US: Nationwide including DC, PR and GU and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BERMUDA, BOSNIA HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHINA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERMANY, GREECE, HONG Kong, HUNGARY, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, KAZAKHSTAN, REPUBLIC OF KOREA, KUWAIT, LEBANON, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PERU, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, and VENEZUELA.

Timeline

Recall initiated
2015-08-31
Posted by FDA
2015-11-10
Terminated
2018-09-07
Status

Source: openFDA Device Recall endpoint. Recall record ID #140519. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.