Product
MYLA MASTER DVD 3.2 CLI DL380 Product Usage: MYLA is a computer application (Middleware) based on Web 2.0 technology which: Interfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s)).
- FDA product code
- JQP — Calculator/Data Processing Module, For Clinical Use
- Device class
- Class 1
- Medical specialty
- Clinical Chemistry
- Affected lot / code info
- REF # 415358, Serial number/Software version V3.2 CLI
Why it was recalled
MYLA software connected to a Laboratory Information System (LIS) and a VITEK 2 system has the potential to link a test result to an incorrect patient record with the same specimen ID.
Root cause (FDA determination)
Software design
Action the firm took
bioMerieux sent an Urgent Product Correction Notice letter dated November10, 2015 via FedEx to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local bioMrieux Clinical Customer Service organization at (800) 682-2666.
Recalling firm
- Firm
- bioMerieux, Inc.
- Address
- 100 Rodolphe St, Durham, North Carolina 27712-9402
Distribution
- Distribution pattern
- Worldwide Distribution - US Nationwide in the states of: AK, AL, AR, AZ, CA, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and countries of: Foreign accounts: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, France, Germany, Greece, Hungary, India, Italy, Japan, Korea, Netherlands, Poland, Portugal, Russia, Serbia, Singapore, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom, Angola, Puerto Rico, Republic of Belarus, Republic of Belarus, Benin, Bosnia, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroon, Chad, Democratic Republic of Congo, Costa Rica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Gabon, Guam, Guatemala, Guatemala, Guinea, Honduras, Indonesia, Iraq, Israel, Ivory Coast, Kazakhstan, Kenya, Kenya, Kuwait, Laos, Latvia, Lebanon, Libya, Lithuania, Malaysia, Malaysia, Mali, Portugal, Myanmar, Managua, Niger, Nigeria, Palestine, Philippines, Philippines, Qatar, Kingdom of Saudi Arabia, Senegal, Serbia, Slovakia, Togo, Tunisia, Ukraine, UAE, Vietnam, Vietnam, Ivory Coast and Zimbabwe.
Timeline
- Recall initiated
- 2015-11-03
- Terminated
- 2017-12-04
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #141769. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.