#141821

Product

1 Step 12 Panel Cup, Item No. NBCA-12M-W

FDA product code
JXMEnzyme Immunoassay, Benzodiazepine
Device class
Class 2
Medical specialty
Clinical Toxicology
510(k) numbers
K052115, K113046, K113501
Affected lot / code info
143107 151657 153155 153529

Why it was recalled

Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

A Medical Device Labeling Correction letter was sent on 11/16/15 to distributors and end users to inform them that Ameditechs Drugs of Abuse Tests have shown reduced reactivity for up to two (2) of the nineteen (19) BZO compounds for which specificity information is provided in the product insert. The letter informs the customers that the particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide. Customers with questions are instructed to contact Ameditech at (844) 616-4453 or email at amt.fieldaction@alere.com.

Recalling firm

Firm
Ameditech Inc
Address
9940 Mesa Rim Rd, San Diego, California 92121-2910

Distribution

Distribution pattern
Nationwide Distribution.

Timeline

Recall initiated
2015-11-16
Terminated
2016-11-09