FDA Device Recalls

Recalls /

#142053

Product

Reagent Management System (RMS) Dimension EXL with LM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension clinical chemistry system. The RMS doubles the available storage of refrigerated Flex reagent cartridges on the Dimension system from 44 to 88. The RMS loads, hydrates, and automatically removes Flex reagent cartridges, allowing the Dimension system to process samples without interruption.

FDA product code
JJEAnalyzer, Chemistry (Photometric, Discrete), For Clinical Use
Device class
Class 1
Medical specialty
Clinical Chemistry
510(k) numbers
K130276
Affected lot / code info
Dimension EXL with LM Catalog Numbers 778021.911 (Domestic)/ 778021.931 (Intl). All serial numbers.

Why it was recalled

Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located directly behind the RMS Waste Container.

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

Siemens sent an Urgent Medical Device Correction (UMDC # DI-16-01.A.US) and Urgent Field Safety Notice (UFSN # DI-16-01.A.OUS) letters dated November 2015 to all affected customers. The letters identified the affected product, problem and actions to be taken. For questions contact your local Siemens technical support representative.

Recalling firm

Firm
Siemens Healthcare Diagnostics Inc.
Address
101 Silvermine Rd, Brookfield, Connecticut 06804-2047

Distribution

Distribution pattern
US Nationwide Distribution: FL, NY, SC, TX, LA, AZ, WI, IA, IL, OH, AL, AK, NH, OK, MD, CA, NM, MS, KY, NJ, MN, GA, CO, MA, SD, TN, MO, PA, MT, SC, NE, KS, VA, AL, NV, WY, OR, PR, MO, OK, GU, HI, AR, ME, VT, CT, NC and the countries of Canada and Mexico.

Timeline

Recall initiated
2015-11-10
Posted by FDA
2015-12-22
Terminated
2017-08-23
Status

Source: openFDA Device Recall endpoint. Recall record ID #142053. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.