Product

BD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x 3.0 mL (Lavender) BD Hemogard Additive: K2EDTA (spray dried), 5.4mg Sterile IVD BD, Made in USA BD Vacutainer tubes are used to collect, transport and process blood in a closed system. These tubes are used for testing serum, plasma or whole blood in the clinical laboratory. The BD Vacutainer Plus K2EDTA Tubes may be used for routine immunohematology testing and blood donor screening.

FDA product code
JKATubes, Vials, Systems, Serum Separators, Blood Collection
Device class
Class 2
Medical specialty
Clinical Chemistry
510(k) numbers
K981013
Affected lot / code info
Catalog Reference #368054; Lot numbers 5089827, 5210514, 5154646, 5210511.  Catalog Reference #367856; Lot number 5064666

Why it was recalled

Some of the Blood Collection Tubes in these lots were manufactured with insufficient EDTA additive. An insufficient EDTA additive may lead to erroneous platelet results, specimen recollection, or a delay of treatment.

Root cause (FDA determination)

Process change control

Action the firm took

The firm, BD, issued an "URGENT PRODUCT RECALL" notifications/Recall Response Forms to their customers dated 12/3/2015 via email and/or UPS 2nd Day Mail. The notification identified the reason for the recall; how to identify affected product; and actions to be taken. The customers were instructed customers to immediately review their inventory; complete the enclosed Recall Response Form and fax (1-866-873-0312 it to BD or email it to bd4964@stericycle.com; and return all affected products with the completed Recall Response Form following instructions on the enclosed packing instruction, even if you do not have any affected lot in your inventory. If customers have any questions or require assistance with the return of the recalled product and/or availability of replacement product, they were instructed to contact 1-888-628-0732 between 8AM and 5 PM ET Monday through Friday.

Recalling firm

Firm
Becton Dickinson & Company
Address
1 Becton Dr, Franklin Lakes, New Jersey 07417-1815

Distribution

Distribution pattern
Worldwide Distribution: US (Nationwide) and the countries of: European Union, Peoples Republic of China, Japan and Republic of Korea.

Timeline

Recall initiated
2015-12-03
Posted by FDA
2015-12-31
Terminated
2018-10-19
Status

Source: openFDA Device Recall endpoint. Recall record ID #142385. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.