FDA Device Recalls

Recalls /

#143462

Product

Metasul Head 40, 12/14, SZ M/0 total hip prosthesis - metal femoral heads Noninflammatory degenerative joint disease (NIDJD) including avascular necrosis, osteoarthritis, post-traumatic arthritis and congenital hip dysplasia and inflammatory joint disease (IJD) e.g. rheumatoid arthritis if bone quality is adequate. Failed previous surgery where pain, deformity, or dysfunction persists. Revision of previously failed hip arthroplasty. Total hip replacements may be considered for younger patients if any unequivocal indication outweighs the risks associated with the age of the patient and modified demands regarding activity and hip joint loading are assured. This includes severely handicapped patients with multiple joint involvement, for whom an immediate need of hip mobility leads to an expectation of significant improvement in the quality of their lives. The system is intended for use either with or without bone cement in total hip arthroplasty."

FDA product code
KWAProsthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)
Device class
Class 3
Medical specialty
Orthopedic
510(k) numbers
K091973
Affected lot / code info
Item number: 00-8770-040-02 Lots: 2564467 2573371 2625382 2631721 2655075 2655076 2685143

Why it was recalled

LDPE bag containing the implant adheres to the highly polished implant surface.

Root cause (FDA determination)

Process control

Action the firm took

Zimmer Biomet sent an Urgent Medical Device Recall dated January 11, 2016, to all affected customers via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with questions or concerns were instructed to contact customer service at 1-800-348-2759 or contact their Zimmer Biomet Sales Representative. For questions regarding this recall call 800-447-5633.

Recalling firm

Firm
Zimmer Biomet, Inc.
Address
1800 W Center St, Warsaw, Indiana 46580-2304

Distribution

Distribution pattern
Worldwide Distribution - US (nationwide) and Internationally to ARGENTINA AUSTRALIA AUSTRIA BARBADOS BELGIUM BERMUDA BOLIVIA BRAZIL CARIBBEAN CHILE CHINA COLOMBIA COST RICA DENMARK DOMINICAN REPUBLIC DORAL FL ECUADOR EL SALVADOR ENGLAND FRANCE GERMANY GUATEMALA HONDURAS HONG KONG INDIA ISRAEL ITALY JAMAICA JAPAN KOREA MALAYSIA MEXICO NEW ZEALAND NICARAGUA PANAMA PERU SINGAPORE SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND VENEZUELA WEST INDIES Canary Islands Croatia Cyprus Czech Republic Canada

Timeline

Recall initiated
2016-01-11
Posted by FDA
2016-02-22
Terminated
2017-05-23
Status