FDA Device Recalls

Recalls /

#143782

Product

LightSpeed Plus (4-slice MDAS LightSpeed 4.0 CT Scanner System - The LightSpeed 4.0 CT Scanner System is indicated for head and whole body X-ray Computed Tomography applications. PET Discovery ST system - The GE Discovery ST system is intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images. The Discovery ST is to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions. This can assist in the evaluation, diagnosis, staging, restaging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. This device can also assist in radiotherapy planning. The Discovery ST system can also be used as a stand-alone head and whole body multi-slice computed tomography (CT) diagnostic imaging system. LightSpeed Pro16 - The Lightspeed 5.0 CT Scanner System is indicated for head and whole body X-ray computed tomography applications. 16 Slice BGO PET/CT System - The GE Discovery XR and XA Systems are intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images. The Discovery XR and XA are to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions. This can assist in the evaluation, diagnosis, staging, restaging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. These devices can also assist in radiotherapy planning. The Discovery XR and XA can also be used as stand-alone head and whole body multislice computed tomography (CT) diagnostic imaging systems. 6 Slice BGO PET/CT System - The GE Discovery LS System is intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images. The Discovery LS is to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions. This can assist in the evaluation, diagnosis, staging, restaging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. This device can also assist in radiotherapy planning. The Discovery LS can also be used as a stand-alone head and whole body multislice computed tomography (CT) diagnostic imaging system.

FDA product code
KPSSystem, Tomography, Computed, Emission
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K040172
Affected lot / code info
Mfg Lot or Serial # System ID 00000002493PT5 0850450029

Why it was recalled

GE Healthcare has updated the Service Manuals for a number of CT and PET CT Scanners to update an incorrect torque value.

Root cause (FDA determination)

Labeling design

Action the firm took

GE Healthcare sent an Urgent Medical Device Correction letter " GEHC Ref# 25467-ABT or 25467" dated February 12, 2016, to all affected consignees . The letter was addressed to Hospital Administrators / Risk Managers, Managers of Radiology/Cardiology, & Radiologist/Cardiologist. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. Consignees were instructed to ensure that the organization that service their equipment uses the latest version of the Service Manual to maintain the safe and proper performance of their medical device. For questions call in the United States: 1-800-437-1171 or your local GE Healthcare Service Representative. For questions regarding this recall call 262-513-4122.

Recalling firm

Firm
GE Medical Systems, LLC
Address
3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615

Distribution

Distribution pattern
Worldwide Distribution - US including all states in including PR except ND, VT & WY. Foreign Distribution: ALBANIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOSNIA & HERZEGOVINA, BRAZIL, BULGARIA, BURKINA FASO, CAMERON, CANADA, CHILE, CHINA, COLOMBIA, DOMINICAN REPUBLIC, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE,HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRA N, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, REPUBLIC OF KOREA, KUWAIT, LATVIA, LEBANON, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, LUXEMBOURG, MALAYSIA, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PERU, PHILIPPINES, POLAND, PORTUGAL, REUNION, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TUNISIA, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA .

Timeline

Recall initiated
2016-02-10
Terminated
2016-09-21
Status

Source: openFDA Device Recall endpoint. Recall record ID #143782. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.