Product

TC-PLUS(TM), Primary Fixed Tibial Component VKS Right, SIZE 8 RIGHT, STERILE R, REF 12 000 0012, S&N 7500267, Smith & Nephew Orthopaedics AG Product Usage: Tibial knee prosthesis

FDA product code
JWHProsthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K072858
Affected lot / code info
Batch number C1507672

Why it was recalled

The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being unable to seat the polyethylene (PE) Insert on the affected VKS tibial base plate during surgery.

Root cause (FDA determination)

Process control

Action the firm took

The firm, Smith & Nephew, issued an "URGENT FIELD SAFETY NOTICE: Medical Device Field Safety Corrective Action/Recall" letter dated 3/10/2016 to its local QA/RA personnel/customers in the affected countries by email on 03/11/2016. The letter described the product, problem and actions to be taken. The customers were instructed to locate and quarantine affected unused devices immediately; return quarantined product to your national Smith & Nephew agency/distributor; complete and return the Return Slip via fax to your national Smith & Nephew agency/distributor; and pass on this safety information to all those who need to be aware in your organization. The local QA/RA staff will then notify end customers according to local procedures. In Germany as well as in Switzerland, sales representatives will inform the affected customers in-person and retrieve affected devices from their stock. If you have any questions, call 978-749-1330 or email: joe.metzger@smith-nephew.com

Recalling firm

Firm
Smith & Nephew, Inc.
Address
1450 E Brooks Rd, Memphis, Tennessee 38116-1804

Distribution

Distribution pattern
International Distribution only in countries of: Germany and Switzerland.

Timeline

Recall initiated
2016-03-11
Posted by FDA
2016-04-21
Terminated
2017-04-10
Status

Source: openFDA Device Recall endpoint. Recall record ID #144577. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.