FDA Device Recalls

Recalls /

#145894

Product

Shuttle Select Slip-Cath Catheter, Intravascular, Diagnostic The catheters are intended for use in angiographic procedures by physicians trained and experienced in angiographic techniques

FDA product code
DQOCatheter, Intravascular, Diagnostic
Device class
Class 2
Medical specialty
Cardiovascular
510(k) numbers
K122937
Affected lot / code info
All Lots. Catalog prefixes/suffixes SCBR5.5-/-SHTL SCBR6.5-/-SHTL  Catalog/product number; Global/order number SCBR5.5-35-125-P-NS-H1-SHTL G31129 SCBR5.5-35-125-P-NS-JB1-SHTL G31130 SCBR5.5-35-125-P-NS-JB2-SHTL G36361 SCBR5.5-35-125-P-NS-MPA-SHTL G23697 SCBR5.5-35-125-P-NS-SIM1-SHTL G23699 SCBR5.5-35-125-P-NS-SIM2-SHTL G36360 SCBR5.5-35-125-P-NS-VERT-SHTL G23698 SCBR5.5-35-125-P-NS-VTK-SHTL G36359 SCBR5.5-35-135-P-NS-ANG-SHTL G48154 SCBR5.5-35-150-P-NS-0-SHTL G23705 SCBR5.5-35-150-P-NS-ANG-SHTL G23704 SCBR5.5-35-75-P-NS-ANG-SHTL G48152 SCBR6.5-35-125-P-NS-H1-SHTL G31134 SCBR6.5-35-125-P-NS-JB1-SHTL G31135 SCBR6.5-35-125-P-NS-JB2-SHTL G36364 SCBR6.5-35-125-P-NS-SIM1-SHTL G23701 SCBR6.5-35-125-P-NS-SIM2-SHTL G36363 SCBR6.5-35-125-P-NS-VERT-SHTL G23700 SCBR6.5-35-125-P-NS-VTK-SHTL G36362

Why it was recalled

Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.

Root cause (FDA determination)

Unknown/Undetermined by firm

Action the firm took

Cook Inc. sent an URGENT: MEDICAL DEVICE RECALL NOTIFICATION letter dated April 15, 2016, to all affected customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with any medical questions or concerns, were instructed to contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235. For information regarding the recall, please contact Stericycle Expert Solutions at 866.201.9067.

Recalling firm

Firm
Cook Inc.
Address
750 N Daniels Way, Bloomington, Indiana 47404-9120

Distribution

Distribution pattern
Worldwide Distribution - US (Nationwide) and Internationally to Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Myanmar, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, "Palestinian Territory, Occupied", Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, U.S. Virgin Islands, Yugoslavia.

Timeline

Recall initiated
2016-04-15
Posted by FDA
2016-08-22
Terminated
2017-08-04
Status

Source: openFDA Device Recall endpoint. Recall record ID #145894. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.