FDA Device Recalls

Recalls /

#146800

Product

Standard Abutment Dental implants

FDA product code
NHAAbutment, Implant, Dental, Endosseous
Device class
Class 2
Medical specialty
Dental
510(k) numbers
K072642
Affected lot / code info
Item: AB200, AB300, AB400, AB500 and AB700.

Why it was recalled

Pouches may not have been sealed during packing.

Root cause (FDA determination)

Packaging

Action the firm took

Biomet sent an Urgent Device Recall Notice to all affected consignees the week of November 30, 2015, communicating the issue and customer responsibilities. International consignees were notified via traceable courier or email week of November 30, 2015. Customers were instructed to return affected product by calling BIOMET at 1-800-342-5454 to obtain a RMA number to return products. Customers were also asked to complete the Business Reply Form and fax to 561-514-6316. For questions regarding this recall call 561-776-6700.

Recalling firm

Firm
Biomet 3i, LLC
Address
4555 Riverside Dr, Palm Beach Gardens, Florida 33410-4200

Distribution

Distribution pattern
Worldwide Distribution Nationwide andHawaii, Guam, Puerto Rico and Virgin Islands (U.S.) . FOREIGN CONSIGNEES: Albania, Andorra, Argentina, Australia, Austria, Barbados, Belgium, Bolivia, Brazil, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Denmark, Dominican Republic, France, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Jamaica, Japan, Korea (South), Kuwait, Lebanon, Liechtenstein, Lithuania, Martinique, Mexico, Mongolia, Netherlands, Northern Mariana Islands, Paraguay, Peru, Poland, Portugal, Qatar, Reunion, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad and Tabago, Turkey, United Arab Emirates, Uruguay and Viet Nam.

Timeline

Recall initiated
2015-11-25
Posted by FDA
2016-08-12
Terminated
2019-07-26
Status