Recalls / —
—#147470
Product
Depuy Synthes Trauma- Trochanteric Fixation Nail Advanced (TFNA) Blade/Screw Guide Sleeve, Surgical Instrument Part Number ID 03.037.017, Class I exempt Lot Numbers: 9066399; 9066435; 9409790; 9519308; 9620447; 9301183; 9409792; 9519303; 9641077; 9301184; 9409791; 9542998; 9641078; 9327510; 9422885; 9554330; 9651850; 9340629; 9351396; 9554329; 9604307; 9347642; 9428332; 9533341; 9661610; 9339162; 9439269; 9574931; 9734868; 9347641; 9485532; 9569618; 9704087; 9351397; 9485534; 9569617; 9734876; 9359540; 9494322; 9590744; 9661608; 9359339; 9501935; 9590747; 9371095; 9505477; 9604306 These devices are specific to the Trochanteric Fixation Nail Advanced (TFNA) system. The TFNA Proximal Femoral Nailing System is for the treatment of intramedullary fixation of proximal femoral fractures
- FDA product code
- FZX — Guide, Surgical, Instrument
- Device class
- Class 1
- Medical specialty
- General, Plastic Surgery
- Affected lot / code info
- Catalog ID: 03.037.017 Lot Numbers: 9066399; 9066435; 9409790; 9519308; 9620447; 9301183; 9409792; 9519303; 9641077; 9301184; 9409791; 9542998; 9641078; 9327510; 9422885; 9554330; 9651850; 9340629; 9351396; 9554329; 9604307; 9347642; 9428332; 9533341; 9661610; 9339162; 9439269; 9574931; 9734868; 9347641; 9485532; 9569618; 9704087; 9351397; 9485534; 9569617; 9734876; 9359540; 9494322; 9590744; 9661608; 9359339; 9501935; 9590747; 9371095; 9505477; 9604306 Manufactured Date : 30Jul2014 -01Jul2015
Why it was recalled
DePuy Synthes is initiating a voluntary medical device recall due to Blade/Screw Guide Sleeves (Part # 03.037.017) may have been manufactured with a slightly larger than specified outer diameter which could prevent the sleeve from going through the TFNA aiming arm properly, which may cause surgical delay.
Root cause (FDA determination)
Nonconforming Material/Component
Action the firm took
DePuy Synthes sent an Urgent Notice Medical Device Recall letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to conduct functional testing to inspect the instrument to determine if the Guide Sleeve is able to fit through the aiming arm properly. Customers were asked to only return the device if the product does not pass. Customers with questions were instructed to call 610-719-5450 or contact their DePuy Synthes Sales Consultant.
Recalling firm
- Firm
- Synthes (USA) Products LLC
- Address
- 1301 Goshen Pkwy, West Chester, Pennsylvania 19380-5986
Distribution
- Distribution pattern
- Nationwide Distribution to AK,AL,AR,AZ,CA,CO,CT,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,PR,RI,SC,SD,TN,TX,UT,VA,VT,WA,WI,WV, and WY
Timeline
- Recall initiated
- 2016-06-07
- Posted by FDA
- 2016-07-15
- Terminated
- 2017-04-25