FDA Device Recalls

Recalls /

#147873

Product

OER-Pro Endoscope Reprocessor, Endoscope washer/disinfector

FDA product code
FEBAccessories, Cleaning, For Endoscope
Device class
Class 2
Medical specialty
Gastroenterology, Urology
510(k) numbers
K103264
Affected lot / code info
Model: OER-Pro, All serial numbers affected

Why it was recalled

New warning requiring Users to conduct duodenoscope precleaning and manual cleaning even when using an Automated Endoscope Reprocessor (AER) that has instructions that may indicate a user could forego certain steps in precleaning and manual cleaning of the endoscopes.

Root cause (FDA determination)

Under Investigation by firm

Action the firm took

Olympus America Inc. (OAI) mailed a letter to customers informing them that they are issuing an updated Operation Manual for the OER-Pro Automated Endoscope Reprocessor. The following actions were asked to be taken: -Olympus has discontinued previously distributed copies of the OER-Pro Operation Manual. - Inspect your inventory of Operation Manuals and discard any existing inventory of OER-Pro Operation Manuals. - Implement use of the enclosed OER-Pro Operation Manual, which contains the new Warning on pages 73 and 80. - Ensure all reprocessing personnel are completely knowledgeable and thoroughly trained on the requirement to perform precleaning and manual cleaning of Olympus duodenoscopes prior to placing the duodenoscopes in the OER-Pro for high level disinfection. - Additional copies of the new OER-Pro Operation Manual can be obtained by contacting our Technical Assistance Center at 1-800-848-9024, option 1, or by indicating on the enclosed questionnaire. Additional Operation Manuals will be mailed to your facility - Please indicate on the enclosed questionnaire that you have received this notification. Fax the completed form to (484) 896-7128.

Recalling firm

Firm
Olympus Corporation of the Americas
Address
3500 Corporate Pkwy PO Box 610, Center Valley, Pennsylvania 18034-0610

Distribution

Distribution pattern
Distributed to: AK,AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,RI,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY

Timeline

Recall initiated
2016-06-23
Posted by FDA
2016-10-19
Terminated
2017-01-26
Status

Source: openFDA Device Recall endpoint. Recall record ID #147873. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.