FDA Device Recalls

Recalls /

#148055

Product

Stryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock, Non-Sterile, Used to complement or replace the standard instruments used during the implantation of hip, knee, trauma and upper extremities implants

FDA product code
MBHProsthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K123486
Affected lot / code info
Catalog No: I-K3254TA00 Lot Nos: F6E13243, F6L13864, F6M14078, F6S14667, F6S14831, F6W15444, F6W15406, F7C15943, F7H16099 and F7L16994

Why it was recalled

It was reported that the secondary locking mechanism, and its corresponding components, disassociated from the Specialty Triathlon Tibial Alignment Handle with Secondary Lock Assembly during surgery. Upon further investigation, it was discovered that the weld, which was intended to hold the secondary locking mechanism together, did not meet the weld size specified on the engineering drawing.

Root cause (FDA determination)

Process control

Action the firm took

Stryker sent an Urgent Product Recall letter dated June 24, 2016, with a Business Reply Form attached to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to inform users of the Urgent Product Recall letter and forward the notice to all those individuals who need to be aware within their organization. Customers were instructed to return all affected product at their location to: Stryker C/O Stericycle 2670 Executive Dr., Suite A Indianapolis, IN 46241 Customers were also instructed to complete and sign the enclosed Business Reply Form and fax it to 888-912-7352 or email to Stericycle at strykerortho4582@stericycle.com. Customers with questions were advised to call (201) 831-6693. For questions regarding this recall call 201-831-5000.

Recalling firm

Firm
Stryker Howmedica Osteonics Corp.
Address
325 Corporate Dr, Mahwah, New Jersey 07430-2006

Distribution

Distribution pattern
Nationwide Distribution to FL, IN, LA, NC, PA, TN and UT

Timeline

Recall initiated
2016-06-23
Terminated
2018-08-01
Status

Source: openFDA Device Recall endpoint. Recall record ID #148055. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.