Recalls / —
—#148281
Product
ARTIS ZEE AND ZEEGO, system, x-ray, angiographic AXIOM Artis, Artis zee / zeego and Artis Q/ Q.zen are a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis, Artis zee / zeego and Artis Q/ Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis, Artis zee / zeego and Artis Q/ Q.zen systems can also support the acquisition of position triggered imaging for spatial data synthesis.
- FDA product code
- IZI — System, X-Ray, Angiographic
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K073290
- Affected lot / code info
- Model Numbers:10094135, 10094137, 10094139, 10094141, 10094142, 10094143, 10280959, 10848280, 10848281, 10848282, 10848283, 10848353, 10848354, 10848355, 5904433, 5904441, 5904466, 5904649, 5904656, 5917054, 7007755, 7008605, 7412807, 7413078, 7555357, 7555365, 7555373, 7727717, 7728350, 7728392
Why it was recalled
The possibility exists for contaminants, in the form of biomass, to develop in the cooling system of Artis systems resulting in damage to the pump system.
Root cause (FDA determination)
Under Investigation by firm
Action the firm took
Siemens has implemented a Customer Safety advisory Notice to all affected customer via Update Instruction AX007/15/S. Additional, a corrective action will be implemented via Update Instructions AX006/15/S and AX003/16/S.
Recalling firm
- Firm
- Siemens Medical Solutions USA, Inc
- Address
- 40 Liberty Blvd, Malvern, Pennsylvania 19355-1418
Distribution
- Distribution pattern
- Nationwide Distribution to: AK,AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,PR,RI,SC,SD,TN,TX,UT,VA,WA,WI,WV,WY, Puerto Rico and Guam
Timeline
- Recall initiated
- 2016-07-13
- Terminated
- 2017-11-08
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #148281. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.