Product

LifeCare PCA 3, PCA Serial List Number 12384 Allows for clinician administration or self-administration of analgesic medications.

FDA product code
MEAPump, Infusion, Pca
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K042800
Affected lot / code info
Serial Number 12921204  12922976  12924008  12925861  12925924  12925998  12928528  12928807  12929016  12929126  12920432  12920820  12920856  12920915  12921404  12921413  12922304  12922976  12925861  12925998  12927215  12929016  12931044  12931141  12931155  12931192  12931244  12931342  12931361  12931413  12931610  12931639  12931649  12931911  12921480  12921487  12923723  12924326  12924350  12924546  12925259  12925447  12926315  12927938  12928075  12928614  12930418  12930709  12930890  12931441  12931760  12931911  12932487  12932717  12920754  12920791  12920897  12921063  12921613  12922062  12923441  12924161  12926859  12927756  12928775  12929346  12929438  12931942  12920112

Why it was recalled

Screw rotation error occurring when the pump detects a mismatch between the number of steps given to the motor driving delivery and the number of steps the motor actually took. When this malfunction occurs, the pump will alarm and infusion stops.

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

Hospira will contract with Stericycle, Inc. in Indianapolis, IN to send letter and a reply form to Hospira, Inc. Receipt of the letters will be confirmed by UPS tracking. Customers will be requested to return a reply form to Stericycle indicating the number of pumps on hand and that they have received the letter and notified their users. In addition, letters will be sent by email to ECRI, NRAC, Noblis and ISMP for notification of their members and posting on their websites. The letter will also be posted to the Hospira website. Letters will be sent to Hospira customers of record who purchased the devices directly from Hospira. Customers will be directed to notify their consignees if they further distributed the product and confirm they have done so. Telephone attempts will be made to contact customers that do not respond in order to confirm receipt of the letter. Final effectiveness will be done as the devices are upgraded with new plastic motors.

Recalling firm

Firm
Hospira Inc.
Address
275 N Field Dr, Lake Forest, Illinois 60045-2579

Distribution

Distribution pattern
United States (Nationwide) Distribution including Puerto Rico and American Virgin Islands.

Timeline

Recall initiated
2013-03-08
Terminated
2016-11-07