FDA Device Recalls

Recalls /

#148604

Product

TIBIAL TRAY , SIZE 5, REF 90-SRK-200500, QTY: 1, STERILE R, Responsive Orthopedics, total knee replacement

FDA product code
JWHProsthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K150496
Affected lot / code info
P239175

Why it was recalled

The alpha total knee replacement components that may have not been passivated in accordance with the requirements of the job traveler.

Root cause (FDA determination)

Employee error

Action the firm took

On 07/18/2016, the firm initiated the correction by verbally notifying (by visit and telephone) three of four (3 of 4) implanting surgeons; the fourth implanting surgeon was verbally notified on 07/25/2016. The firm followed with a letter to all four (4) surgeons on 07/28/2016. Medtronic recommended that the condition of the patients receiving the subject part be assessed for any indications of premature failure during normal post-operative follow-up visits.

Recalling firm

Firm
Medtronic Sofamor Danek USA Inc
Address
1800 Pyramid Pl, Memphis, Tennessee 38132-1703

Distribution

Distribution pattern
AR, MN,MA

Timeline

Recall initiated
2016-07-18
Terminated
2016-11-22
Status