Product

Siemens RAPIDPoint 405 Blood Gas Analyzer nBili Siemens Material Number (SMN): 10282093, 10310464, 10314817, 1031 7193, 10318999, 10320055, 10321238, 10322347, 10328278, 10328302, 10336784

FDA product code
CHLElectrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Device class
Class 2
Medical specialty
Clinical Chemistry
510(k) numbers
K110277
Affected lot / code info
All Serial Numbers

Why it was recalled

There is a potential for the first and/or last name of one patient to be printed with Patient ID and result data from a different patient, even though those fields have been turned Off for the system

Root cause (FDA determination)

Software design

Action the firm took

Siemens Healthcare issued Urgent Field Safety Notice to (US) customers via Federal Express. . All affected customers outside the US provided a copy of an Urgent Field Safety Notice via hard copy, e-mail, and/or fax as determined by each countries local regulations and procedures. This notice informs customers of the issue, Actions to be Taken by the Customer and to return the response form. Questions contact your Siemens Customer Care Center or your local Siemens technical support representative

Recalling firm

Firm
Siemens Healthcare Diagnostics Inc
Address
2 Edgewater Dr, Norwood, Massachusetts 02062-4637

Distribution

Distribution pattern
Nationwide Foreign: Albania Algeria Angola Argentina Armenia Australia Austria Bahrain Bangladesh Belarus Belgium Bosnia Herzeg. Botswana Brazil Brunei Bulgaria Canada Chile Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland France Fren.Polynesia Georgia Germany Greece Hong Kong Hungary Iceland India Indonesia Iraq Ireland Israel Italy Ivory Coast Japan Kazakhstan Kenya Kuwait Latvia Lebanon Lesotho Lithuania Luxembourg Macedonia Malaysia Mexico Netherlands New Caledonia New Zealand Norway P.R. China Peru Philippines Poland Portugal Qatar Rep. of Yemen Republic Korea Romania Russian Fed. Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan Thailand Tunisia Turkey Turkmenistan U.A.E. United Kingdom Uruguay Vatikancity Vietnam

Timeline

Recall initiated
2016-08-11
Terminated
2020-07-16