FDA Device Recalls

Recalls /

#149119

Product

Stryker Orthopaedics Patella Assembly Instrument, REF 3182-1000, Non-Sterile, NDC (01) 07613327215854 (10) The Scorpio Patella Assembly Instrument is used during total knee replacement surgery for the implantation of Scorpio Patella implant components. The device functions to fully seat the implant onto the bone and secure it in position while the surgeon removes any excess cement and when cement polymerization occurs.

FDA product code
MBHProsthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K033972
Affected lot / code info
Catalog No. 3182-1000 Lot #'s: RD8E100, RD8A054TT, RD8A054T, RD8E099M, RD8E099A, RD8E099, RD8A054A, RD8E099D, RD8A053A, RD8A054, RD8A053T, RD7K062L, RD8A052A, RD8A052E, RD8A052, RD8A053, RD7E128D, RD5T290H, RD7E128T, RD5S262D, RD5T290W, RD7K062E, RD5S262X, RD7E128Y, RD7A076MA, RD6W102MA, RD7K062A, RD7K062, RD7E128H, RD7A076A, RD7E128A, RS7E128, RD6W103D, RD6V108J, RD6V108K, RD6V108H, RD7A076H, RD6V112M, RD7A074D, RD7A076J, RD6V112H, RD6W102m, RD5H286A, RD2V022, RD2K049, RD6W103, RD5H286J, RD2T037A, RD7A076M, RD6S136A, RD5H286, RD2V031C, RD5N361T, RD6V112A, RD4M068J, RD2V021C, RD7A076, RD6V112, RD4E226W, RD2V021, RD7A074H, RD6W102, RD4S220W, RD2V023X2, RD6W103A, RD6S136M, RD4M069D, RD2V023X3, RD6V108, RD5W148M, RD4S220, RD2V023, RD7A074, RD5M237K, RD4S220A, RD2V031A, RD7A074A, RD5T290X, RD4S220A1, RD2V023X1, RD6W103M, RD5T290A, RD4M069, RD2V031, RD5K116A, RD6S137A, RD4L128X1, RD2V031X2, RD5M236T, RD5W148A, RD4L128V, RD2V031X1, RD5K116T, RD6S137, RD4E226J, RD2N030K, RD3V002A, RD6S136, RD4M068< RD2T037, RD5M237, RD5W148X, RD4L128, RD2T037X, RD5K116N, RD5W148, RD4E226M, RD2N030E, RD5K116, RD5T290M, RD4E226L1, RD2N030S, RD5M236, RD5T290E, RD4C230L, RD2N030X, RD5H286L, RD5T290, RD4E226L, RD2N030XX, RD3T155L, RD5T290D, RD4E226, RD2N030, RD2V021E, RD5N361A, RD4C162L, RD2K068W, RD2V022W, RD5S305J, RD4C162, RD2N030X1, RD2V021S, RD5S305M, RD4C230, RD2M019E, RD2V022K, RD5N361M, RD4C230X, RD2M039, RD2V022E, RD5S305, RD3V002P, RD2M019H, RD2V021H, RD5S262A, RD3V002X1, RD2K049H, RD2V023E, RD5S262M1, RD3V002X2, RD2M003H, RD2T037W, RD5S262M, RD3V002D1, RD2M038H, RD2K068X1, RD5S262J, RD3V002, RD2K068B, RD2K068, RD5S262, RD3V002D, RD2M003, RD2K049R, RD5N361H, RD3V206D, RD2M038, RD2K049E, RD5M259M, RD3T160L1, RD2M019, RD2K017E, RD5N316, RD3V206X1, RD2K068S, RD2H056, RD5H286D, RD3T155L1, RD2K056S, RD2K056, RD5M259T, RD3T160L3, RD2K065S, RD2K065, RD5M236H, RD3T160, RD2K049S, RD2K017, RD5M259, RD3T160L2, RD2K068E, RD5M237L, RD3T160L, RD2H056E, RD5K116J, RD3V206, RD2K065E, RD4S220D, RD3T155 and RD2K056E

Why it was recalled

Upon failure of the press-fit, the clamping sub components and pins could disassociate from the Scorpio Patella Assembly Instrument.

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

Stryker sent an "Urgent Medical Device Recall Notification" Letter and Recall Business Reply Form dated August 1, 2016 to their affected customers via email and also via UPS on August 1, 2016. Customers are asked to quarantine the product subject to this recall. Branches/Agencies are asked to return all affected products available at their location to: Stryker C/O Stericycle, Attn: RA 2016-098 - Event 8296. Hospital/Branches: Customers are asked to complete and sign the Business Reply Form and fax a copy within 5 business days to: 877-272-6141 or email to Stericycle strykerortho8296@stericycle.com. Any questions, customers can call (201) 831-6693.

Recalling firm

Firm
Stryker Howmedica Osteonics Corp.
Address
325 Corporate Dr, Mahwah, New Jersey 07430-2006

Distribution

Distribution pattern
US Nationwide and Internationally

Timeline

Recall initiated
2016-08-01
Posted by FDA
2016-09-13
Terminated
2017-06-08
Status

Source: openFDA Device Recall endpoint. Recall record ID #149119. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.