FDA Device Recalls

Recalls /

#149418

Product

Visualase Visualization Stylets PN 020-2301. The individual components within the Visualase Body Accessory Kit are packaged in Tyvek pouches. The packaged devices are then placed into a cardboard box which is labeled with the Body Accessory Kit label. Labeled sterile. The visualization stylet is constructed of titanium and used during Visualase laser ablation procedures by inserting it into intracorporeal catheter lumens for improved catheter stiffness and MRI visualization during procedures

FDA product code
GAHStylet, Surgical, General & Plastic Surgery
Device class
Class 1
Medical specialty
General, Plastic Surgery
Affected lot / code info
Lot number 151481

Why it was recalled

Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls and/or bends toward the field.

Root cause (FDA determination)

Mixed-up of materials/components

Action the firm took

Medtronic sent an Urgent Medical Device Recall letter dated September 2, 2016, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to examine their stock for the affected product and if any are found contact Medtronic to arrange for a no charge replacement. Customers were asked to complete the Customer Confirmation form attached to the letter and return it to Medtronic at RS.NavFCA@medtronic.com or by Fax to 651-367-7075. Customers with questions were instructed to contact Medtronic Technical Services at 1-800-595-9709. For questions regarding this recall call 720-890-3200.

Recalling firm

Firm
Medtronic Navigation, Inc.
Address
826 Coal Creek Cir, Louisville, Colorado 80027-9710

Distribution

Distribution pattern
Worldwide Distribution - US including CA, GA, FL, and TX. Internationally to Australia.

Timeline

Recall initiated
2016-09-01
Posted by FDA
2016-10-28
Terminated
2017-03-27
Status

Source: openFDA Device Recall endpoint. Recall record ID #149418. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: Visualase Visualization Stylets PN 020-2301. The individual components within the Visualase Body Acc · FDA Device Recalls