Product
cobas 8000 Modular Analyzer Series c8000, Chemistry (Photometric, Discrete) for clinical use Modular e602: Immunoassay Analyzer Product Usage: The VS II system is a stand-alone, computer-controlled instrument designed to aliquot patient sample from a primary tube to a barcoded secondary tube for processing on an off-line analyzer.
- FDA product code
- JJE — Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Device class
- Class 1
- Medical specialty
- Clinical Chemistry
- 510(k) numbers
- K100853
- Affected lot / code info
- Not applicable
Why it was recalled
Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
Root cause (FDA determination)
Device Design
Action the firm took
On 7/15/2016, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, an updated cleaning procedure, and instructions for responding to the formal recall notification. Please contact the Roche Support Network Customer Support Center at 1-800-428-2336 if you have questions about the information contained in this UMDC. On 10/5/2016, URGENT MEDICAL DEVICE CORRECTION UPDATE notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. The notification updated the 7/15/16 notification to include additional affected systems and to provide updated cleaning instructions for all systems via the enclosed Operators Manual addendum.
Recalling firm
- Firm
- Roche Diagnostics Operations, Inc.
- Address
- 9115 Hague Rd, Indianapolis, Indiana 46256-1025
Distribution
- Distribution pattern
- US Nationwide Distribution
Timeline
- Recall initiated
- 2016-07-15
- Terminated
- 2017-06-05
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #149603. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.