FDA Device Recalls

Recalls /

#150867

Product

SOMATOM Force, System x-ray, tomography, computed

FDA product code
JAKSystem, X-Ray, Tomography, Computed
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K133589
Affected lot / code info
Model Number 10742326, Serial Numbers: 75708,75709,75607,75617,75550,75526 75512,75567,75536,75711,75514,75600,75563,75510,75454,75540,75571,75500 75527,75475,75535,75481,75482,75710,75460,75476,75588,75493,75458,75528 75467,75602,75487,75464,75478,75604,75513,75532,75555,75524,75515,75559 75610,75501,75570,75616,75576,75605

Why it was recalled

Siemens is providing software update version VA50A_SP3 to address the software bugs that were identified through normal field monitoring and the Global Complaint Handling Process. Correction for the problems are as follows: 1. Correction to volumetric misrepresentations of high contrast objects when using ADMIRE. 2. Correction to highly sporadic scan aborts due to temporarily tube currents at 0mA. 3. Correction to missing Microsoft Hotfixes (MS16-001, MS15-135, MS15-088, MS15-048). 4. Correction to highly sporadic image artifacts caused by error signals from high voltage chain at extremely high tube currents. 5. Correction to highly sporadic incomplete image reconstruction due to temporary slow hard disk write speed. 6. Correction to highly sporadic scan aborts due to temporarily high disk load of system disk. 7. Correction to highly sporadic image artifacts generated by scatter correction in case of tube arcings.

Root cause (FDA determination)

Under Investigation by firm

Action the firm took

Siemens sent a Customer Information Letter dated November 2, 2016, to customers affected by the recall to inform them of a release of a software update that will provide bug-fixes to improve system performance. Customers will be contacted to arrange a date for onsite installation. For further questions, please call (610) 448-6471.

Recalling firm

Firm
Siemens Medical Solutions USA, Inc
Address
40 Liberty Blvd, Malvern, Pennsylvania 19355-1418

Distribution

Distribution pattern
Nationwide Distribution.

Timeline

Recall initiated
2016-10-20
Posted by FDA
2016-12-02
Terminated
2017-05-30
Status

Source: openFDA Device Recall endpoint. Recall record ID #150867. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.