Recalls / —
—#150872
Product
Atrium Medical Ocean Drains as follows: 2002-000 Ocean single drain, with in-line connectors and suction control stopcock 2002-040 Ocean single drain, in-line connector and suction control stopcock, Pedi connector in bag on back of the drain 2002-050 Ocean single drain with in-line connector and suction control stopcock, Pre-connected 2550 ATS Bag 2002-100 Ocean single drain with in-line connector, no suction control stopcock 2002-300 Ocean single drain with suction control stopcock, no in-line connector 2002-400 Ocean single drain, no in-line connector or suction control stopcock, Water ampoule Product Usage: To evacuate air and/or fluid from the chest cavity or mediastinum. To help re-establish lung expansion and restore breathing dynamics. To facilitate postoperative collection and reinfusion of autologous blood from the patient s pleural cavity or mediastinal area.
- FDA product code
- PAD — Chest Drainage Kit
- Device class
- Class 2
- Medical specialty
- General Hospital
- Affected lot / code info
- All Lot Numbers With Expiration Date Prior To October 2019
Why it was recalled
Outer Packaging is not sterile
Root cause (FDA determination)
Package design/selection
Action the firm took
Atrium Medical issued recall letter on 11/29/16 via two-day express mail delivery advising users: of the problem, updated instructions for use; the new Warning Label applied to the packaging; and requesting to post the warning label near all Atrium chest drain product inventory. A Field Notification Reply Form acknowledging receipt of the Field Notification was requested to be completed and returned. Questions contact Atrium/Maquet/Getinge chest drainage representative or Atrium Customer Service at 1-800-528-7486 (603-880-1433), Monday through Friday between 8:00 am and 5:00 pm (EST).
Recalling firm
- Firm
- Atrium Medical Corporation
- Address
- 5 Wentworth Dr, Hudson, New Hampshire 03051-4929
Distribution
- Distribution pattern
- Worldwide Distribution - US Nationwide and the countries of Algeria, Australia, Austria, Belgium, Brazil, Canada, Chile, China, China, Columbia, Ecuador, Egypt, Finland, France, Germany, Great Britain, Greece, Guatemala, Hong Kong, Ireland, Italy, Kuwait, Lebanon, Libya, Mexico, Netherlands, New Zealand, Pakistan, Panama, Peru, Poland, Portugal, Puerto Rico, Saudi Arabia, South Korea, Spain, Sweden, Thailand, Tunisia, and Venezuela,
Timeline
- Recall initiated
- 2016-11-29
- Terminated
- 2020-09-29
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #150872. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.