FDA Device Recalls

Recalls /

#150896

Product

Artis zee/zeego, Angiographic x-ray system Product Usage: The Artis systems are a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee / zeego and Q/ Q.zen include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee / zeego and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.

FDA product code
OWBInterventional Fluoroscopic X-Ray System
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K141574
Affected lot / code info
Model Numbers: 10094135, 10094137, 10094139, 10094141, 10280959

Why it was recalled

The possibility exists that the semi-automatic focus switch may not work as intended when pressing the footswitch multiple times.

Root cause (FDA determination)

Software in the Use Environment

Action the firm took

Siemens mailed a Customer Safety Advisory Notice to affected customers to inform them of a corrective action they plan to perform to prevent a possible hazard to persons and equipment. The notice identified the associated risks, underlying issues, impact on system operation and actions to be taken.

Recalling firm

Firm
Siemens Medical Solutions USA, Inc
Address
40 Liberty Blvd, Malvern, Pennsylvania 19355-1418

Distribution

Distribution pattern
US Nationwide Distribution.

Timeline

Recall initiated
2016-11-01
Posted by FDA
2016-11-17
Terminated
2017-06-21
Status

Source: openFDA Device Recall endpoint. Recall record ID #150896. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: Artis zee/zeego, Angiographic x-ray system Product Usage: The Artis systems are a family of dedicat · FDA Device Recalls