Recalls / —
—#151090
Product
Maquet Oasis Dry Suction Water Seal Chest Drain, Single Collection w/AC Reference: 3600-100 Intended Use: To evacuate air and/or fluid from the chest cavity or mediastinum. " To help re-establish lung expansion and restore breathing dynamics. " To facilitate postoperative collection and reinfusion of autologous blood from the patients pleural cavity or mediastinal area
- FDA product code
- CAC — Apparatus, Autotransfusion
- Device class
- Class 2
- Medical specialty
- Anesthesiology
- 510(k) numbers
- K043140
- Affected lot / code info
- Lot Number: 242909, 242910, 242912, 242914, 242915
Why it was recalled
Chest drains manual vent valve located at the top of affected chest drains could potentially not re-seal after being depressed.
Root cause (FDA determination)
Process control
Action the firm took
Maquet issued a Field action correction notification sent to customers on 12/23/16 via Fed'x for delivery on Tuesday Dec.27, 2016.who have received affected product and describes the failure mode. Accounts to complete the enclosed Field Notification Reply Form to acknowledge receipt of the notification. Return the Field Notification Reply Form to the following e-mail address: drainsmvv.us@getinge.com or you may fax the form to 1-603-386-6590. Questions, contact your local Atrium/Maquet/Getinge chest drainage representative or Atrium Customer Service at 1-800-528-7486 (603-880-1433), Monday through Friday between 8:00 am and 5:00 pm (EST). UPDATE: On 1/16/17, Maquet issued recall letters that identified correct lot numbers and products in the letter titles: UPDATE-UPDATED AFFECTED DEVICES. Two letters issued one to customers who received the Dec 23 letter and one to new customers who who did not receive the Dec. 23 letter.
Recalling firm
- Firm
- Atrium Medical Corporation
- Address
- 5 Wentworth Dr, Hudson, New Hampshire 03051-4929
Distribution
- Distribution pattern
- Worldwide Distribution - US (nationwide) and Internationally to Canada and Mexico
Timeline
- Recall initiated
- 2016-12-23
- Terminated
- 2020-11-03
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #151090. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.