Product
Allura Xper FD20 R8.2 722028 is intended for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and Electrophysiology(EP).
- FDA product code
- OWB — Interventional Fluoroscopic X-Ray System
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K133292
- Affected lot / code info
- 1000
Why it was recalled
Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled.
Root cause (FDA determination)
Process control
Action the firm took
Customers were notified of the recall via letter sent on 12/1/16. The letter explains the issue and the action that Philips plans to take.
Recalling firm
- Firm
- Philips Electronics North America Corporation
- Address
- 3000 Minuteman Rd, Andover, Massachusetts 01810-1032
Distribution
- Distribution pattern
- US: WA OUS: China, Spain, Korea, Republic of, Germany
Timeline
- Recall initiated
- 2016-12-01
- Posted by FDA
- 2017-01-20
- Terminated
- 2017-08-10