FDA Device Recalls

Recalls /

#151909

Product

ADVANTX LCLP+

FDA product code
IZISystem, X-Ray, Angiographic
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K974367
Affected lot / code info
Mfg. Lot or Serial # System ID 00000047915YY4 415540ADVLCLP 00000063721YY5 909865LCLP Not Available 916734XR 00000063736YY3 MCTHKGMN 00000023830YY3 912633MCL1 00000023761YY0 912633MCL1 00000128484M03 312CCHGECATH 00000128484MO3 312CCHGECATH Not Available 507255CC101 Not Available LIJ015064 00000026160YY2 210428VCATH2 00000063805YY6 BPEP Not Available 414C4862 00000022150YY7 YX0738

Why it was recalled

GE Healthcare is conducting a recall due to the potential fall of aged CRT monitors - Prestige 2, Prestilix, Prestige VH, Prestige SI, RFX/SFX, Legacy, Precision 500D, ADVANTX and Innova 2000 systems.

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

Consignees were sent on 11/21/2016 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# FMI 10913 dated November 21, 2016. The letter was addressed to Director of Radiology, Risk Manager/Hospital Administrator, and Director of Biomedical Engineering. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Recalling firm

Firm
GE Healthcare, LLC
Address
3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615

Distribution

Distribution pattern
US: AL, AR, AZ, CA, CO, CT, DC, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT,NC, NE NH, NJ, NM, NY, NV, OH, OK, PA, RI, SC, SD, TN, TX,, VA, WA, WO, WY, WV. OUS: Algeria Armenia Australia Austria Brazil Canada Chile China Denmark Egypt France Germany Guatemala Honduras Hong Kong Hungary Iceland India Ireland Italy Japan Korea Libya Mexico Netherlands Netherlands Peru Poland Portugal Russia Saudi Arabia Serbia Sierra Leone Slovakia Spain Taiwan Thailand Turkey United Arab Emirates United Kingdom Venezuela Vietnam

Timeline

Recall initiated
2016-11-21
Terminated
2017-09-26
Status

Source: openFDA Device Recall endpoint. Recall record ID #151909. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.