#151910

Product

ADVANTX LCN+

FDA product code
IZISystem, X-Ray, Angiographic
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K974367
Affected lot / code info
Mfg. Lot or Serial # System ID 00000089619MO1 205592MLC 00000089632MO4 205592MLC 00000052196YY3 601459LC 00000023729YY7 910343NHC1 00000031441YY9 603668CLC3 00000034211YY3 603668CLC3 00000034271YY7 973429CATH Not Available 215576LC2 Not Available 610250SC Not Available 0910062022 Not Available 0910062022 00000023777YY6 1025 00000026231YY1 3919-1 00000023841YY0 93162 00000023841YY0 93162 Not Available GOLD000540

Why it was recalled

GE Healthcare is conducting a recall due to the potential fall of aged CRT monitors - Prestige 2, Prestilix, Prestige VH, Prestige SI, RFX/SFX, Legacy, Precision 500D, ADVANTX and Innova 2000 systems.

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

Consignees were sent on 11/21/2016 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# FMI 10913 dated November 21, 2016. The letter was addressed to Director of Radiology, Risk Manager/Hospital Administrator, and Director of Biomedical Engineering. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Recalling firm

Firm
GE Healthcare, LLC
Address
3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615

Distribution

Distribution pattern
US: AL, AR, AZ, CA, CO, CT, DC, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT,NC, NE NH, NJ, NM, NY, NV, OH, OK, PA, RI, SC, SD, TN, TX,, VA, WA, WO, WY, WV. OUS: Algeria Armenia Australia Austria Brazil Canada Chile China Denmark Egypt France Germany Guatemala Honduras Hong Kong Hungary Iceland India Ireland Italy Japan Korea Libya Mexico Netherlands Netherlands Peru Poland Portugal Russia Saudi Arabia Serbia Sierra Leone Slovakia Spain Taiwan Thailand Turkey United Arab Emirates United Kingdom Venezuela Vietnam

Timeline

Recall initiated
2016-11-21
Terminated
2017-09-26
Status