#151917

Product

Prestilix

FDA product code
LLZSystem, Image Processing, Radiological
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K924310
Affected lot / code info
Mfg. Lot or Serial # System ID Not Available 1301 Not Available 318783ARF Not Available 412488F1 Not Available AP3001XR10 Not Available 902567GI Not Available C4211004 00000048037YY6 PC7281XR02 Not Available PC0731XR01 Not Available A5816801 00000022153YY1 C5304806 00000023739YY6 PL1020RX01 Not Available RU1146RX01

Why it was recalled

GE Healthcare is conducting a recall due to the potential fall of aged CRT monitors - Prestige 2, Prestilix, Prestige VH, Prestige SI, RFX/SFX, Legacy, Precision 500D, ADVANTX and Innova 2000 systems.

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

Consignees were sent on 11/21/2016 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# FMI 10913 dated November 21, 2016. The letter was addressed to Director of Radiology, Risk Manager/Hospital Administrator, and Director of Biomedical Engineering. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Recalling firm

Firm
GE Healthcare, LLC
Address
3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615

Distribution

Distribution pattern
US: AL, AR, AZ, CA, CO, CT, DC, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT,NC, NE NH, NJ, NM, NY, NV, OH, OK, PA, RI, SC, SD, TN, TX,, VA, WA, WO, WY, WV. OUS: Algeria Armenia Australia Austria Brazil Canada Chile China Denmark Egypt France Germany Guatemala Honduras Hong Kong Hungary Iceland India Ireland Italy Japan Korea Libya Mexico Netherlands Netherlands Peru Poland Portugal Russia Saudi Arabia Serbia Sierra Leone Slovakia Spain Taiwan Thailand Turkey United Arab Emirates United Kingdom Venezuela Vietnam

Timeline

Recall initiated
2016-11-21
Terminated
2017-09-26