Recalls / —
—#152271
Product
Sheridan/CF, Cuffed Tracheal Tube, Catalog # 5-10114 and Sheridan/HVT Cuffed Tracheal Tube, Catalog # 5-10315 A tracheal tube is inserted into a patients mouth or nose for airway management
- FDA product code
- BTR — Tube, Tracheal (W/Wo Connector)
- Device class
- Class 2
- Medical specialty
- Anesthesiology
- 510(k) numbers
- K822082
- Affected lot / code info
- Lots 73A1500461 & 73D1400082
Why it was recalled
Labeling inconsistency - units labeled with a 7mm tube may contain a size 6.5mm tube, and units labeled 7.5mm tube may contain a size 5.5mm tube.
Root cause (FDA determination)
Other
Action the firm took
Teleflex sent an Medical Device Recall Notification letter dated October 18, 2016, to all affected customers. Customers were advised to immediately discontinue use and to quarantine product. The letter to distributors requested a sub-recall. Both letters included an acknowledgement form to be returned. Customers with questions were instructed to contact their local representative or Customer Service at 1-800-246-6990.
Recalling firm
- Firm
- Teleflex Medical
- Address
- 2917 Weck Dr, Research Triangle Park, North Carolina 27709-0186
Distribution
- Distribution pattern
- Worldwide Distribution - US including AL, AZ, ARK, CA, CO, CT, FL, GA. HI, KS, KY, IL, IN, IA, LA, MD, MA, MI, MN, MS, MO, MT, NE, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV and Internationally to Canada, Chile, & Japan
Timeline
- Recall initiated
- 2016-10-18
- Posted by FDA
- 2017-02-09
- Terminated
- 2017-08-25
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #152271. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.