#152271

Product

Sheridan/CF, Cuffed Tracheal Tube, Catalog # 5-10114 and Sheridan/HVT Cuffed Tracheal Tube, Catalog # 5-10315 A tracheal tube is inserted into a patients mouth or nose for airway management

FDA product code
BTRTube, Tracheal (W/Wo Connector)
Device class
Class 2
Medical specialty
Anesthesiology
510(k) numbers
K822082
Affected lot / code info
Lots 73A1500461 & 73D1400082

Why it was recalled

Labeling inconsistency - units labeled with a 7mm tube may contain a size 6.5mm tube, and units labeled 7.5mm tube may contain a size 5.5mm tube.

Root cause (FDA determination)

Other

Action the firm took

Teleflex sent an Medical Device Recall Notification letter dated October 18, 2016, to all affected customers. Customers were advised to immediately discontinue use and to quarantine product. The letter to distributors requested a sub-recall. Both letters included an acknowledgement form to be returned. Customers with questions were instructed to contact their local representative or Customer Service at 1-800-246-6990.

Recalling firm

Firm
Teleflex Medical
Address
2917 Weck Dr, Research Triangle Park, North Carolina 27709-0186

Distribution

Distribution pattern
Worldwide Distribution - US including AL, AZ, ARK, CA, CO, CT, FL, GA. HI, KS, KY, IL, IN, IA, LA, MD, MA, MI, MN, MS, MO, MT, NE, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV and Internationally to Canada, Chile, & Japan

Timeline

Recall initiated
2016-10-18
Posted by FDA
2017-02-09
Terminated
2017-08-25
Status