FDA Device Recalls

Recalls /

#152774

Product

14 mm Distraction Pin, Exagg Qty: 1 per pack; STERILE; Manufactured for: Medline Industries, Inc., Mundelein, IL 60060 USA Distraction Pin is designed for Anterior Cervical Fusion Procedures. Use with vertebral body distraction instruments. It is intended as a temporary fixation screw utilized during cervical spine procedures.

FDA product code
LXHOrthopedic Manual Surgical Instrument
Device class
Class 1
Medical specialty
Orthopedic
Affected lot / code info
Item# MDS9091414A; Lots #136190

Why it was recalled

Product's non-conformity involves the integrity of the seal in the sterile packaging. It is possible that the seal on the sterile packaging has been compromised resulting in a loss of sterility of the medical device contained within.

Root cause (FDA determination)

Process control

Action the firm took

Medline Industries sent an Immediate Action Required letter dated December 14, 2016, to all affected customers with response forms via US mail, notifying them of the recall. Customers were instructed to quarantine affected product and return it to the firm. The product will be repackaged and sterilized. Customers with questions were instructed to call 866-359-1704. For questions regarding this recall call 847-643-3245.

Recalling firm

Firm
MEDLINE INDUSTRIES INC
Address
3 Lakes Dr, Northfield, Illinois 60093-2753

Distribution

Distribution pattern
Nationwide Distribution

Timeline

Recall initiated
2016-12-14
Terminated
2019-07-09
Status