Recalls / —
—#152812
Product
Breast Augmentation Pack, part number AMS1786(A Breast Augmentation Pack, part number AMS2790(A Breast Augmentation Pack, part number AMS3324(B Breast Augmentation Pack, part number AMS4382 Breast Augmentation Pack, part number AMS4382(A Breast Augmentation Pack, part number AMS4382(B Breast Augmentation Pack, part number AMS4392 Breast Augmentation Pack, part number PSS3430
- FDA product code
- FSY — Light, Surgical, Ceiling Mounted
- Device class
- Class 2
- Medical specialty
- General, Plastic Surgery
- Affected lot / code info
- Lot number and expiration date 100634, 9/23/2018 83241, 6/20/2017 83757, 6/1/2017 84174, 1/14/2017 84249, 1/20/2017 84496, 6/21/2017 84521, 3/19/2017 84904, 4/28/2017 84910, 1/25/2017 85071, 11/30/2017 85126, 6/28/2017 85126, 6/28/2017 85162, 1/17/2017 85565, 1/24/2017 85634, 6/11/2017 85820, 3/25/2017 85831, 6/14/2017 85985, 7/15/2017 86062, 4/21/2017 86141, 3/19/2017 86656, 7/10/2017 86797, 2/5/2018 86862, 6/18/2017 87093, 6/26/2017 87110, 8/22/2017 87747, 3/21/2017 88532, 2/11/2018 89014, 9/9/2017 89155, 12/22/2017 89244, 12/13/2017 89569, 12/3/2017 89770, 5/3/2017 89932, 7/7/2017 94601, 5/16/2017 95184, 7/29/2017 95185, 5/12/2017 95224, 5/1/2017 95265, 3/20/2018 95623, 9/22/2017 95793, 12/1/2017 96299, 5/22/2019 96563, 3/28/2018 96987, 5/21/2017 97563, 12/22/2017 97564, 3/30/2018 97902, 2/9/2018 98436, 3/7/2018 98445, 12/25/2018 98900, 2/28/2018 98937, 9/12/2018 98980, 2/3/2018 99258, 3/8/2018
Why it was recalled
The Medtronic COVER LIGHT HANDLE DISPOSABLE GLOVES of the custom Aligned Medical Solutions (AMS) kits were recalled because gloves may contain splits or holes compromising sterility.
Root cause (FDA determination)
Nonconforming Material/Component
Action the firm took
AMS notified their consignees by Email, Fax, and Phone on 11/11/2017. AMS sent the notification letter, dated January 11, 2017, and provided yellow stickers "Recall Notice" to customers. They are advised to over-label the affected AMS kits at their facilities. Customers were instructed to place sticker in such a way that it would be immediately noticed but should not cover up or conceal any pertinent information such as Pack Number, Pack Name, Lot Number or Expiration Date on a kit. Customers with questions can call Vicki Davis at 321-527-7714 or send email to vdavis@alignedmedicalsolutions.com.
Recalling firm
- Firm
- Windstone Medical Packaging, Inc.
- Address
- 1602 4th Ave N, Billings, Montana 59101-1521
Distribution
- Distribution pattern
- Distributed in AZ, CA, CO, CT, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, ND, NE, NJ, NV, OH, PA, TX, UT, VA, and WA.
Timeline
- Recall initiated
- 2017-01-11
- Posted by FDA
- 2017-02-28
- Terminated
- 2017-05-03
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #152812. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.