FDA Device Recalls

Recalls /

#152890

Product

Lower Extremity Pack, part number AMS4669

FDA product code
FSYLight, Surgical, Ceiling Mounted
Device class
Class 2
Medical specialty
General, Plastic Surgery
Affected lot / code info
Lot number and expiration date  AMS4669, 85376 AMS4669, 86279 AMS4669, 87657 AMS4669, 88434 AMS4669, 95690 AMS4669, 96189 AMS4669, 96190 AMS4669, 99735 AMS4669, 100446

Why it was recalled

The Medtronic COVER LIGHT HANDLE DISPOSABLE GLOVES of the custom Aligned Medical Solutions (AMS) kits were recalled because gloves may contain splits or holes compromising sterility.

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

AMS notified their consignees by Email, Fax, and Phone on 11/11/2017. AMS sent the notification letter, dated January 11, 2017, and provided yellow stickers "Recall Notice" to customers. They are advised to over-label the affected AMS kits at their facilities. Customers were instructed to place sticker in such a way that it would be immediately noticed but should not cover up or conceal any pertinent information such as Pack Number, Pack Name, Lot Number or Expiration Date on a kit. Customers with questions can call Vicki Davis at 321-527-7714 or send email to vdavis@alignedmedicalsolutions.com.

Recalling firm

Firm
Windstone Medical Packaging, Inc.
Address
1602 4th Ave N, Billings, Montana 59101-1521

Distribution

Distribution pattern
Distributed in AZ, CA, CO, CT, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, ND, NE, NJ, NV, OH, PA, TX, UT, VA, and WA.

Timeline

Recall initiated
2017-01-11
Posted by FDA
2017-02-28
Terminated
2017-05-03
Status

Source: openFDA Device Recall endpoint. Recall record ID #152890. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.