Recalls / —
—#153924
Product
Poly Component Trial, CR, packaged in the following sizes and configurations: a) Poly Component Trial - CR 3x10, REF 90-SRK-160310 b) Poly Component Trial - CR 3x12, REF 90-SRK-160312 c) Poly Component Trial - CR 3x14, REF 90-SRK-160314 d) Poly Component Trial - CR 4x10, REF 90-SRK-160410 e) Poly Component Trial - CR 4x12, REF 90-SRK-160412 f) Poly Component Trial - CR 4x14, REF 90-SRK-160414 g) Poly Component Trial - CR 5x10,REF 90-SRK-160510 h) Poly Component Trial - CR 5x12, REF 90-SRK-160512 i) Poly Component Trial - CR 5x14, REF 90-SRK-160514 j) Poly Component Trial - CR 6x10, REF 90-SRK-160610 k) Poly Component Trial - CR 6x12, REF 90-SRK-160612 l) Poly Component Trial - CR 6x14, REF 90-SRK-160614 The Responsive Orthopedics (RO) Total Knee Arthroplasty (TKA) System is intended to restore alignment, stability, range of motion, and alleviate pain by replacing the articulating surfaces of the knee joint
- FDA product code
- JWH — Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Device class
- Class 2
- Medical specialty
- Orthopedic
- 510(k) numbers
- K150496
- Affected lot / code info
- Lot Numbers: a) REF 90-SRK-160310, TU53730-01 b) REF 90-SRK-160312, TU53731-01 c) REF 90-SRK-160314, TU53732-01 d) REF 90-SRK-160410, TU53730-02 e) REF 90-SRK-160412, TU53731-02 f) REF 90-SRK-160414, TU53732-02 g) REF 90-SRK-160510, TU53730-03 h) REF 90-SRK-160512, TU53731-03 i) REF 90-SRK-160514, TU53732-03 j) REF 90-SRK-160610, TU53730-04 k) REF 90-SRK-160612, TU53731-04 l) REF 90-SRK-160614, TU53732-04
Why it was recalled
The dovetails of poly trials, Beta 2.0, were observed as either cracked or broken.
Root cause (FDA determination)
Under Investigation by firm
Action the firm took
Medtronic initiated their recall on 20 February 2017, by letter delivered by Fed Ex . The letter was addressed to the Risk/Materials Manager and stated that a Medtronic Clinical Specialist would coordinate the removal of the Responsive Orthopedics Total Knee Arthroplasty (TKA) System instrument sets from the premises. Customers with questions were instructed to contact the Medtronic Post-Market Quality Team at 901-399-3197.
Recalling firm
- Firm
- Medtronic Sofamor Danek USA Inc
- Address
- 1800 Pyramid Pl, Memphis, Tennessee 38132-1703
Distribution
- Distribution pattern
- US distribution to AR, MA
Timeline
- Recall initiated
- 2017-02-20
- Terminated
- 2017-10-27
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #153924. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.