FDA Device Recalls

Recalls /

#153926

Product

Poly Component Trial, PS, packaged in the following sizes and configurations: a) Poly Component Trial - PS 3x10, REF 90-SRK-180310 b) Poly Component Trial - PS 3x12, REF 90-SRK-180312 c) Poly Component Trial - PS 3x14, REF 90-SRK-180314 d) Poly Component Trial - PS 4x10, REF 90-SRK-180410 e) Poly Component Trial - PS 4x12, REF 90-SRK-180412 f) Poly Component Trial - PS 4x14, REF 90-SRK-180414 g) Poly Component Trial - PS 5x10, REF 90-SRK-180510 h) Poly Component Trial - PS 5x12, REF 90-SRK-180512 i) Poly Component Trial - PS 5x14, REF 90-SRK-180514 j) Poly Component Trial - PS 6x10, REF 90-SRK-180610 k) Poly Component Trial - PS 6x12, REF 90-SRK-180612 l) Poly Component Trial - PS 6x14, REF 90-SRK-180614 The Responsive Orthopedics (RO) Total Knee Arthroplasty (TKA) System is intended to restore alignment, stability, range of motion, and alleviate pain by replacing the articulating surfaces of the knee joint

FDA product code
JWHProsthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K150496
Affected lot / code info
Lot Numbers: a) REF 90-SRK-180310, TU53738-01 b) REF 90-SRK-180312, TU53739-01 c) REF 90-SRK-180314, TU53740-01 d) REF 90-SRK-180410, TU53738-02 e) REF 90-SRK-180412, TU53739-02 f) REF 90-SRK-180414, TU53740-02 g) REF 90-SRK-180510, TU53738-03 h) REF 90-SRK-180512, TU53739-03 i) REF 90-SRK-180514, TU53740-03 j) REF 90-SRK-180610, TU53738-04 k) REF 90-SRK-180612, TU53739-04 l) REF 90-SRK-180614, TU53740-04

Why it was recalled

The dovetails of poly trials, Beta 2.0, were observed as either cracked or broken.

Root cause (FDA determination)

Under Investigation by firm

Action the firm took

Medtronic initiated their recall on 20 February 2017, by letter delivered by Fed Ex . The letter was addressed to the Risk/Materials Manager and stated that a Medtronic Clinical Specialist would coordinate the removal of the Responsive Orthopedics Total Knee Arthroplasty (TKA) System instrument sets from the premises. Customers with questions were instructed to contact the Medtronic Post-Market Quality Team at 901-399-3197.

Recalling firm

Firm
Medtronic Sofamor Danek USA Inc
Address
1800 Pyramid Pl, Memphis, Tennessee 38132-1703

Distribution

Distribution pattern
US distribution to AR, MA

Timeline

Recall initiated
2017-02-20
Terminated
2017-10-27
Status

Source: openFDA Device Recall endpoint. Recall record ID #153926. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.