Recalls / —
—#154083
Product
CADD Administration Set, 21-7339-24, Rx Only, Sterile EO.
- FDA product code
- FPA — Set, Administration, Intravascular
- Device class
- Class 2
- Medical specialty
- General Hospital
- 510(k) numbers
- K040636
- Affected lot / code info
- 46X817, 46X829, 46X899, 46X828, 46X896, 46X990, 46X831, 46X897, 46X1001.
Why it was recalled
Smiths Medical has become aware that the drawing on the Instructions for Use (IFU) on certain yellow striped CADD Administration Sets (part number 21-7339-24), incorrectly represent the actual filter size of the product. Although the text description on the IFU correctly identifies the product to include a 0.2 micron filter, the illustration on the IFU identifies the filter size as 2.0 microns.
Root cause (FDA determination)
Packaging process control
Action the firm took
Smiths Medical sent an "Urgent Medical Device Field Safety Notice" dated March 13, 2017. The letter was addressed to "Risk Manager". It provided a list of the Affected devices, Risk to Health, Instructions to Distributors & Customers. Advised consignees to notify their customers. There is no need to return the devices to Smiths Medical. Requested consignees to complete the Urgent Medical Device Field Safety Notice (FSN) Response Form and return to smithsmedical3947@stericycle.com. For questions contact Stericycle at 844-265-7407. For further questions regarding this recall , please call (844) 265-7407.
Recalling firm
- Firm
- Smiths Medical ASD Inc.
- Address
- 6000 Nathan Ln N, Minneapolis, Minnesota 55442-1690
Distribution
- Distribution pattern
- US:AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MS, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WV, WY. OUS: Canada
Timeline
- Recall initiated
- 2017-03-17
- Posted by FDA
- 2017-03-30
- Terminated
- 2019-07-03
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #154083. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.