FDA Device Recalls

Recalls /

#154118

Product

STAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da Vinci Xi Greek SRK General and Plastic Surgery: The Intuitive Surgical EndoWrist Stapler 45, Stapler 45 Reloads and other Stapler Accessories are intended to be used with the da Vinci Xi Surgical System (Model IS4000) for resection, transection and/or creation of anastomoses in General, Thoracic, Gynecologic, and Urologic surgery. The device can be used with staple line or tissue butressing material (natural or synthetic).

FDA product code
NAYSystem, Surgical, Computer Controlled Instrument
Device class
Class 2
Medical specialty
Gastroenterology, Urology
510(k) numbers
K140553
Affected lot / code info
Part number: 381387-01

Why it was recalled

Intuitive Surgical has become aware and further investigation has determined that misinterpretation of instructions provided on the IS4000 Stapler Release Kit instruction card may result damage of the tool.

Root cause (FDA determination)

Device Design

Action the firm took

Intuitive sent an Urgent Medical Device Correction (ISIFA2017-02-C) letter dated March 16, 2017 were sent by FED EX to be received on March 17, 2017. Letters advised users of the reason for the notification, risk to health, affected products with part numbers and a copy of Appendix A - proper use of the Stapler Release Kit (SRK). Customers were instructed to distribute a copy of the letter to all users at their facility, follow all instructions using the Instrument Release Kit (IRK) tool to finish manually unclamping the stapler and return the SRK to Intuitive Surgical through the Returned Material Authorization (RMA) process. Customers were also instructed to log into the da Vinci Online Community Field Action resource (US only) to read or complete any request actions related to this issue. In the case where the da Vinci online resource cannot be used (including outside he US), complete the attached Acknowledgement Form and return it via fax to Intuitive Surgical as instructed on the form. Customers were asked to retain a copy of the letter with their Stapler User Manual and the acknowledgement form for their files. If you need further information or support concerning this notification , please contact Intuitive Surgical Customer Service at the numbers listed below: North and South America: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: customersupport-servicesupport@intusurg.com Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 AM to 6 PM CET) or ics@intusurg.com South Korea: 02-3271-3200 (9 AM to 6 PM KSTJ) Japan: 0120-56-5635 or 03-5575-1362 (9 AM to 6 PM JST) For questions regarding this recall call 408-523-2100.

Recalling firm

Firm
Intuitive Surgical, Inc.
Address
1266 Kifer Rd, Bldg 100, Sunnyvale, California 94086-5304

Distribution

Distribution pattern
Worldwide Distribution - US (nationwide) and Internationally to Australia, Austria, Belgium, Canada, Chile, Colombia, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Japan, Luxembourg, Monaco, Netherlands, Norway, Portugal, Puerto Rico, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom and Venezuela

Timeline

Recall initiated
2017-03-15
Terminated
2018-12-10
Status

Source: openFDA Device Recall endpoint. Recall record ID #154118. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.