FDA Device Recalls

Recalls /

#154246

Product

The following systems are only affected if a Laird chiller is installed (11NC: 98960021599): Allura Xper Releases 3 to 8.1: Allura Xper FD10 (C+F). 722003; Allura Xper FD10/10 722005; Allura Xper FD20 722006; Allura Xper FD20/10 and FD20/20 722008; Allura Xper FD10 722010; Allura Xper FD10/10 722011; Allura Xper FD20 722012; Allura Xper FD20 biplane 722013; Allura Xper FD10 OR Table 722014; Allura Xper FD20 OR Table 722015; Allura Xper FD10/10 OR Table 722019; Allura Xper FD20 Biplane OR Table 722020; Allura Xper FD10 OR Table 722022; Allura Xper FD20 OR Table 722023; Allura Xper FD10/10 OR Table 722024; Allura Xper FD20 Biplane OR Table 722025; Allura CV20 722031

FDA product code
OWBInterventional Fluoroscopic X-Ray System
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K031333, K033737, K041949, K102005
Affected lot / code info
Devices with a Laird Chiller.

Why it was recalled

Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.

Root cause (FDA determination)

Under Investigation by firm

Action the firm took

Consignees were notified of the recall via letter sent on 3/22/2017. The letter explains the issue, actions that should be taken by the customer, and actions to be taken by Philips.

Recalling firm

Firm
Philips Electronics North America Corporation
Address
3000 Minuteman Rd, Andover, Massachusetts 01810-1032

Distribution

Distribution pattern
Nationwide, worldwide including Canada

Timeline

Recall initiated
2017-03-22
Terminated
2021-05-17
Status

Source: openFDA Device Recall endpoint. Recall record ID #154246. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.