Recalls / —
—#154710
Product
Arrow VPS Access Kit with Maximal Barrier Precautions for use with 5 Fr. Peripherally inserted Central Venous Catheter The Arrow(r) CVC is indicated to provide short-term (<30 days) central venous access for treatment of diseases or conditions requiring central venous access including, but not limited to: multiple infusions of fluids, medications, or chemotherapy; infusion of fluids that are hypertonic, hyperosmolar, or have divergent pH values; frequent blood sampling or blood/blood component infusions; infusion of incompatible medications; central venous pressure monitoring; lack of usable peripheral IV sites; replacement of multiple peripheral sites for IV access.
- FDA product code
- FOZ — Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
- Device class
- Class 2
- Medical specialty
- General Hospital
- 510(k) numbers
- K002507
- Affected lot / code info
- Material Number ASK-04001-DU7
Why it was recalled
Arrow International initiated the recall due to reports of safety cover disengagement and needlestick injury for the BD Eclipse Needle that are packaged with certain Arrow products.
Root cause (FDA determination)
Under Investigation by firm
Action the firm took
Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.
Recalling firm
- Firm
- Arrow International Inc
- Address
- 2400 Bernville Rd, Reading, Pennsylvania 19605-9607
Distribution
- Distribution pattern
- Nationwide Distribution
Timeline
- Recall initiated
- 2017-03-20
- Terminated
- 2018-12-18
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #154710. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.