Recalls / —
—#154803
Product
Logic Fit Tibial Tamp Head
- FDA product code
- LXH — Orthopedic Manual Surgical Instrument
- Device class
- Class 1
- Medical specialty
- Orthopedic
- Affected lot / code info
- Catalog #: 213-75-01 Lot #'s: 37166-002, 44944-004, 44944-005, 44944-006, 48073-010, 48073-011, 48073-012, 49036-017, 51180-004, 51180-005, 51180-006, 51180-008, 51754-031, 51754-032, 51754-033, 51754-034, 59743-001, 61872-001,and 61872-002.
Why it was recalled
The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to remove the tamp assembly from the bone, i.e., to include a caution statement about the potential for instrument breakage if the tamp handle/guide is misused by impacting in retro-grade.
Root cause (FDA determination)
Error in labeling
Action the firm took
Consignees were sent a Field Advisory Notice dated 5/22/2017 by mail informing them updated surgical techniques are available on Exactech's website (www.exac.com). The notice requested consignees: Extend the information to accounts in possession of the product Complete & return the attached Field Advisory Notice Response Form to Exactech
Recalling firm
- Firm
- Exactech, Inc.
- Address
- 2320 NW 66th Ct, Gainesville, Florida 32653-1630
Distribution
- Distribution pattern
- AL, AZ, CA, CA, CO, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, RI, SC, TN, TX, VA, WA, WI, Hawaii and Puerto Rico Australia, France, Germany, India, Italy, Japan, Korea, Luxembourg, Singapore, Spain, Switzerland, The Netherlands and United Kingdom
Timeline
- Recall initiated
- 2017-05-25
- Posted by FDA
- 2017-06-19
- Terminated
- 2019-07-26
- Status
- —