Product
VITEK 2 Gram Negative Susceptibility card (AST-GN13), REF 22095, 20 cards per carton.
- FDA product code
- LON — System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
- Device class
- Class 2
- Medical specialty
- Microbiology
- Affected lot / code info
- 1100152203 27/Apr/18 1100180103 25/May/18 1100201103 15/Jun/18 1100252203 05/Aug/18 1100258203 11/Aug/18 1100259203 12/Aug/18
Why it was recalled
The integrity of the product container may be compromised thereby exposing product to moisture which could lead to antibiotic degradation (loss of potency).
Root cause (FDA determination)
Packaging
Action the firm took
The firm posted their recall notification on their customer website on 4/20/2017. The firm sent the recall notification letter to their customers via FedEx on 4/21/2017 for delivery 4/24/2017. A follow-up recall notice will be issued on 5/08/2017 which identified 3 additional lots impacted by the recall.
Recalling firm
- Firm
- Biomerieux Inc
- Address
- 595 Anglum Rd, Hazelwood, Missouri 63042-2320
Distribution
- Distribution pattern
- Nationwide distribution in the U.S, including military and government sites. The product was also distributed to Algeria, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia - Herzegovina, Brazil, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroun, Canada, Chad, Chile, China, Colombia, Congo, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Gabon, Gambia, Georgia, Germany, Greece, Grenada, Guatemala, Guinea, Haiti, Honduras, Hungary, India, Indonesia, Iraq, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Uganda, Kuwait, Laos, Latvia, Lebanon, Lithuania, Malaysia, Mali, Malta, Mexico, Mongolia, Morocco, Mozambique, Myanmar, Netherlands, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, and Zimbabwe.
Timeline
- Recall initiated
- 2017-04-20
- Posted by FDA
- 2017-05-22
- Terminated
- 2020-07-22
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #155061. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.