FDA Device Recalls

Recalls /

#155262

Product

VITEK 2 Gram Positive Susceptibility card (AST-GP75), IVD, REF 415670, 20 cards per carton.

FDA product code
LONSystem, Test, Automated, Antimicrobial Susceptibility, Short Incubation
Device class
Class 2
Medical specialty
Microbiology
Affected lot / code info
2750173203 18/May/18 2750179103 24/May/18 2750210203 24/Jun/18 2750222103 06/Jul/18 2750236103 20/Jul/18 2750236123 20/Jul/18 2750249103 02/Aug/18 2750277103 30/Aug/18 2750285103 07/Sep/18

Why it was recalled

The integrity of the product container may be compromised thereby exposing product to moisture which could lead to antibiotic degradation (loss of potency).

Root cause (FDA determination)

Packaging

Action the firm took

The firm posted their recall notification on their customer website on 4/20/2017. The firm sent the recall notification letter to their customers via FedEx on 4/21/2017 for delivery 4/24/2017. A follow-up recall notice will be issued on 5/08/2017 which identified 3 additional lots impacted by the recall.

Recalling firm

Firm
Biomerieux Inc
Address
595 Anglum Rd, Hazelwood, Missouri 63042-2320

Distribution

Distribution pattern
Nationwide distribution in the U.S, including military and government sites. The product was also distributed to Algeria, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia - Herzegovina, Brazil, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroun, Canada, Chad, Chile, China, Colombia, Congo, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Gabon, Gambia, Georgia, Germany, Greece, Grenada, Guatemala, Guinea, Haiti, Honduras, Hungary, India, Indonesia, Iraq, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Uganda, Kuwait, Laos, Latvia, Lebanon, Lithuania, Malaysia, Mali, Malta, Mexico, Mongolia, Morocco, Mozambique, Myanmar, Netherlands, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, and Zimbabwe.

Timeline

Recall initiated
2017-04-20
Posted by FDA
2017-05-22
Terminated
2020-07-22
Status

Source: openFDA Device Recall endpoint. Recall record ID #155262. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.