Product

Arrow VPS Access Kit with Maximal Barrier Precautions for use with 4 Fr. Peripherally Inserted Central Venous Catheter The intended use of the VPS Stylet and Console (VPS System) is to quickly and accurately guide market available peripherally inserted central catheters (PICCs) to the goal location which is to the cavo-atrial junction near the sino-atrial node

FDA product code
FOZCatheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K002507
Affected lot / code info
Material number ASK-04001-DU8

Why it was recalled

Arrow International initiated the recall due to reports of safety cover disengagement and needlestick injury for the BD Eclipse Needle that are packaged with certain Arrow products.

Root cause (FDA determination)

Under Investigation by firm

Action the firm took

Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.

Recalling firm

Firm
Arrow International Inc
Address
2400 Bernville Rd, Reading, Pennsylvania 19605-9607

Distribution

Distribution pattern
Nationwide Distribution

Timeline

Recall initiated
2017-03-20
Terminated
2018-12-18