FDA Device Recalls

Recalls /

#155963

Product

Philips DuraDiagnost X-ray systems: DuraDiagnost (R1.0, Product Number 712210) DuraDiagnost (R2.0, Product Number 712211) DuraDiagnost (R3.0, Product Number 712211)

FDA product code
KPRSystem, X-Ray, Stationary
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K141381
Affected lot / code info
SN120001-SN120003, SN130001-SN130023, SN140001-SN140022, SN140024, SN140027, SN140030-SN140031, SN140033-SN140046, SN140048, SN140050-SN140068, SN150001-SN150004, SN150006, SN150008-SN150011, SN160001, SN160002

Why it was recalled

Tube arm assembly could fall down due to cracked welding joints

Root cause (FDA determination)

Device Design

Action the firm took

Philips Healthcare sent an URGENT - Field Safety Notice Medical Device Correction letter dated May 23, 2017, advising users of the problem and asking facilities to stop using the device until a Philips Engineer can go on-site to perform the FC071200173; contact your local Philips representative if you have any concern or support required on this issue. If you need any further information or support concerning this issue, contact your local Philips representative. For further questions, please call (978) 659-3000.

Recalling firm

Firm
Philips Electronics North America Corporation
Address
3000 Minuteman Rd, Andover, Massachusetts 01810-1032

Distribution

Distribution pattern
Worldwide Distribution - US Distribution to the states of : Ca and Oh., and to the countries of: Australia, Libya, China, Mayotte, Mexico, France, United Kingdom, Germany, Ecuador, Australia, Spain, Switzerland, Macedonia, Uzbekistan, Kuwait, Malaysia, Netherlands, Argentina, Chile, Colombia, Egypt, United Arab Emirates, Saudi Arabia, Hungary, Peru, Poland, Indonesia, Burkina Faso, Kazakhstan, Turkey and Slovenia.

Timeline

Recall initiated
2017-05-31
Terminated
2018-08-24
Status

Source: openFDA Device Recall endpoint. Recall record ID #155963. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.