Recalls / —
—#156271
Product
BD 10mL Syringe Luer-Lok Tip Catalog number 301997 The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection
- FDA product code
- FMF — Syringe, Piston
- Device class
- Class 2
- Medical specialty
- General Hospital
- 510(k) numbers
- K110771
- Affected lot / code info
- 6061746 2021-02 6082982 2021-02 6089821 2021-03 6117858 2021-04 6117863 2021-04 6117889 2021-04 6117891 2021-04 6146962 2021-04 6173697 2021-05 6173767 2021-06 6173779 2021-06 6208614 2021-07
Why it was recalled
BD identified that several lots did not receive the specified dose of irradiation necessary to meet the Sterility Assurance Level (SAL) required.
Root cause (FDA determination)
Process control
Action the firm took
BD sent recall letter/return response form on 9/16/2016.
Recalling firm
- Firm
- Becton Dickinson & Company
- Address
- 1 Becton Dr, Franklin Lakes, New Jersey 07417-1815
Distribution
- Distribution pattern
- Nationwide Distribution
Timeline
- Recall initiated
- 2016-09-16
- Terminated
- 2018-03-20
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #156271. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.