FDA Device Recalls

Recalls /

#156671

Product

ARROW PICC powered by ARROW VPS Stylet The Pressure Injectable PICC is indicated for short or long term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion, power injection of contrast media and allows for central venous pressure monitoring

FDA product code
LJSCatheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K121941
Affected lot / code info
Lot/Batch Number: 23F16M0263

Why it was recalled

There is a possibility that the finished good kits may contain 5 Fr peel-away sheaths instead of the required 4 Fr peel-away sheaths.

Root cause (FDA determination)

Mixed-up of materials/components

Action the firm took

Arrow International sent an Urgent Medical Device Recall Notification letter dated June 1, 2017, and Recall Acknowledgement Form to their domestic consignees via FedEx 2-day. Customers were advised to immediately inspect stock, discontinue use and quarantine any products with the product codes and lot numbers listed, so that the affected products can be returned to Arrow. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow the firm to document the amount of product you have on hand for return. A customer service representative will the customer with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Arrow International. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow the company to document your receipt of the letter. Customers that have any questions can contact their local sales representative or Customer Service at 1-866-246-6990.

Recalling firm

Firm
Arrow International Inc
Address
2400 Bernville Rd, Reading, Pennsylvania 19605-9607

Distribution

Distribution pattern
Nationwide Distribution to AZ, CA, MA, MO, NC, NY, OR & SD

Timeline

Recall initiated
2017-06-01
Terminated
2020-04-13
Status

Source: openFDA Device Recall endpoint. Recall record ID #156671. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.