FDA Device Recalls

Recalls /

#157814

Product

ROSA Surgical Device 2.5.8 It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide. Guidance is based on a pre-operative plan developed with three-dimensional imaging software, and uses fiducial markers or optical registration. The system is intended for use by neurosurgeons to guide standard neurosurgical instruments. It is indicated for any neurosurgical condition in which the use of stereotactic surgery may be appropriate.

FDA product code
HAWNeurological Stereotaxic Instrument
Device class
Class 2
Medical specialty
Neurology
510(k) numbers
K092239, K101791
Affected lot / code info
Serial No. RO10011, RO14033, RO14031, RO14035, RO13027, RO13023, RO10014, RO10009, RO10010, RO14030, RO11016, RO11017, RO12022, RO14029, RO14032, RO08003, RO09004, RO10007, RO10013, RO12018, RO12019, RO13025, RO14034, RO11015, RO13024, RO09005, RO13026, RO12021, and RO12020.

Why it was recalled

Software corrections reactivating the cooperative endoscopy mode.

Root cause (FDA determination)

Software Design Change

Action the firm took

The firm, Zimmer Biomet Inc., sent Field Service Technicians to consignee locations to perform system upgrade. All US units have been corrected. If you any question, please contact Post Market Surveillance and Regulatory Compliance Director at 574-372-4487 or email to: kevin.escapule@zimmberbiomet.com.

Recalling firm

Firm
Zimmer Biomet, Inc.
Address
1800 W Center St, Warsaw, Indiana 46580-2304

Distribution

Distribution pattern
Worldwide distribution: US (nationwide) in states of: AR, GA, MI, OH, and TX; and countries of:Canada, China, France, Germany, Israel, Italy, Russia, Saudi Arabia, and Spain.

Timeline

Recall initiated
2014-06-04
Terminated
2020-06-16
Status

Source: openFDA Device Recall endpoint. Recall record ID #157814. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.