FDA Device Recalls

Recalls /

#157861

Product

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System, Software Version 3.2. The DxH Slidemaker Stainer allows for adaptation of the smear appearance and stain methodology according to user preferences. Blood smears produced by the Slidemaker portion of the DxH Slidemaker Stainer are moved to baskets for transfer to the Stainer portion by a robot arm. Microscopic examination of the stained blood smears can be used to help determine the hematologic status of a patient. A stained blood film: Allows for the differentiation of white blood cells, facilitates the characterization of red blood cells and platelets, and aids in the identification of blood components and cellular abnormalities.

FDA product code
GKZCounter, Differential Cell
Device class
Class 2
Medical specialty
Hematology
510(k) numbers
K140911
Affected lot / code info
P/N A85371AA20; Catalog No. 775222

Why it was recalled

The firm has identified that a fitting in some Slidemaker Stainer units may not completely seal after installation.

Root cause (FDA determination)

Employee error

Action the firm took

Beckman Coulter sent an Urgent Medical Device Recall letter dated June 20, 2017. The notification advised consignees to immediately contact a Beckman Coulter representative if a leak is observed to schedule a service, to inspect reagent supply drawer for spills, and to follow best laboratory practices when using the device (including Personal Protective Equipment). Beckman Coulter indicates that the issue will be resolved with the replacement of the correct fitting on all impacted instruments. Labs were advised to notify all staff of the recall and to retain the notification. They were also instructed to forward the recall notice to all customers, if any product was further distributed, and to complete and return the Customer Response form within ten days of receiving the recall notification. Completed forms are to be sent back to the firm via email to Regulatory.Notifications@beckmancoulter.com. Regulatory Affairs Mail Code 31-B06. Questions can be directed to beckmancoulter.com or 800-526-7694. Consignees outside of the US are advised to contact their Beckman Coulter representative in the case that questions do arise.

Recalling firm

Firm
Beckman Coulter Inc.
Address
250 S Kraemer Blvd, Brea, California 92821-6232

Distribution

Distribution pattern
Worldwide Distribution - USA (nationwide) Distribution to the states of : AL, AR, AZ, CA, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NJ, NV, NY, OH, OR, PA, RI, SC, TN, TX, VA, WA, and WV., and to the countries of : Australia, Brazil, Canada, Chile, China, Czech Republic, Denmark, France, Greece, Hong Kong, India, Israel, Italy, Japan, Morocco, Portugal, South Korea, South Africa, Spain, Taiwan, Turkey, United Kingdom, Vietnam, and Switzerland.

Timeline

Recall initiated
2017-06-20
Terminated
2021-04-21
Status

Source: openFDA Device Recall endpoint. Recall record ID #157861. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.