FDA Device Recalls

Recalls /

#159486

Product

Medtronic Ascenda(R) Intrathecal Catheter, Models: (a) 8780, UDI: 00643169202122, 00643169783027 (b) 8781, UDI: 00643169202139 (c) 8784, UDI: 00643169202153 The Ascenda Intrathecal Catheter kits (models 8780, 8781 and 8784) contain parts of an infusion system that delivers drugs to the intrathecal space. The implanted system consists of an implantable infusion pump and an Ascenda catheter. The catheter connects to the pump at the catheter port and enters the intrathecal space.

FDA product code
LKKPump, Infusion, Implanted, Programmable
Device class
Class 3
Medical specialty
Unknown
PMA numbers
P860004
Affected lot / code info
(a) Model 8780: N740471001, N740482002, N740507001, N741669001, N741918005, N742124002, N742180002, N742214002, N743749002, N747547001, N740471002, N740482003, N740507002, N741669002, N742077001, N742124004, N742180003, N742214003, N743749004, N747547003, N740471003, N740482004, N740507003, N741669003, N742077002, N742124005, N742180004, N742214004, N743749005, N747547004, N740471004, N740482005, N740507004, N741669004, N742077003, N742124006, N742180006, N742214005, N743749006, N747547005, N740471005, N740482006, N740507005, N741669005, N742077004, N742143001, N742184003, N742942001, N743773001, N740471006, N740496001, N740507006, N741669006, N742077005, N742143002, N742184004, N742942002, N743773002, N740476001, N740496003, N740508001, N741677001, N742077006, N742143005, N742184005, N742942003, N743773003, N740476002, N740496004, N740508002, N741677002, N742088001, N742143006, N742184006, N742942004, N743773004, N740476004, N740496005, N740508003, N741677004, N742088002, N742151001, N742197001, N742942005, N743773005, N740476006, N740496006, N740508004, N741677006, N742088003, N742158002, N742197002, N742977001, N743800006, N740477001, N740500001, N740508006, N741688001, N742088004, N742158003, N742197003, N742977002, N744024001, N740477002, N740500003, N740511001, N741688002, N742088005, N742158004, N742197004, N742977003, N744024002, N740477003, N740500004, N740511002, N741688003, N742088006, N742158005, N742197005, N742977004, N744024003, N740477004, N740500005, N740511003, N741688004, N742093002, N742158006, N742197006, N742977005, N744024004, N740477005, N740500006, N740511004, N741688005, N742093004, N742172001, N742211001, N743157002, N744024005, N740477006, N740506001, N740511005, N741688006, N742093005, N742172002, N742211002, N743157003, N744024006, N740479003, N740506002, N740540002, N741918001, N742112003, N742172003, N742211003, N743157004, N747546001, N740479004, N740506003, N740540003, N741918002, N742112004, N742172004, N742211004, N743157005, N747546003, N740479005, N740506004, N740540005, N741918003, N742112006, N742172005, N742211005, N743157006, N747546004, N740482001, N740506006, N740540006, N741918004, N742124001, N742180001, N742214001, N743749001, N747546005   (b) Model 8781: N745579001, N745579002, N745579005, N745579006, N746858001  (c) Model 8784: N745557001, N745557002, N745557003, N745557004, N745557005

Why it was recalled

There is a possibility that specific lots of catheters that have recently been distributed may be at risk for an increased potential for kinking at the proximal end where the catheter connects to the drug infusion pump.

Root cause (FDA determination)

Other

Action the firm took

Medtronic on September 21, 2017, initiated a medical device recall and sent the following via mail (with delivery confirmation) to the consignees. In addition to the mailing, Medtronic field representatives may personally visit the accounts to provide the recall notification, assist the customer as necessary with locating and returning affected product(s), and verify that the customer has or will respond to Medtronic with the Customer Confirmation Form. For further questions, please call (800) 707-0933.

Recalling firm

Firm
Medtronic Neuromodulation
Address
7000 Central Ave Ne, Minneapolis, Minnesota 55432-3568

Distribution

Distribution pattern
Nationwide Distribution

Timeline

Recall initiated
2017-09-21
Posted by FDA
2018-02-06
Terminated
2020-03-20
Status

Source: openFDA Device Recall endpoint. Recall record ID #159486. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.