FDA Device Recalls

Recalls /

#160251

Product

IntelliVue Multi Measurement Server, IntelliVue Multi Measurement Server SLCP, IntelliVue Multi Measurement Server X2, IntelliVue Patient Monitor Models MP2 and MP5 Product Usage: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment.

FDA product code
MHXMonitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Device class
Class 2
Medical specialty
Cardiovascular
510(k) numbers
K150310, K150975
Affected lot / code info
Model #: M300IA/AL, M3002A, M8102A, M8105A with ECO FW revision E.0 1.22 and 12-lead option C#12 Lots #s included on Consignee list  Catalog #: 862442 (M3001A), 866462 (M3001AL), 865039 (M3002A), 865040 (M8102A), 865024 (M8105A)

Why it was recalled

Distorted ST-Segment when using 12-Lead ECG Monitoring with Philips IntelliVue Patient Monitors

Root cause (FDA determination)

Software design

Action the firm took

Philips sent an Urgent Medical Device Correction letter to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that Philips will provide software update and is provding interim instructions for the operator. For questions contact your local Philips representative.

Recalling firm

Firm
Philips Electronics North America Corporation
Address
3000 Minuteman Rd, Andover, Massachusetts 01810-1032

Distribution

Distribution pattern
Worldwide Distribution and US Nationwide

Timeline

Recall initiated
2017-08-23
Status

Source: openFDA Device Recall endpoint. Recall record ID #160251. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: IntelliVue Multi Measurement Server, IntelliVue Multi Measurement Server SLCP, IntelliVue Multi Meas · FDA Device Recalls