Recalls / —
—#160251
Product
IntelliVue Multi Measurement Server, IntelliVue Multi Measurement Server SLCP, IntelliVue Multi Measurement Server X2, IntelliVue Patient Monitor Models MP2 and MP5 Product Usage: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment.
- FDA product code
- MHX — Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K150310, K150975
- Affected lot / code info
- Model #: M300IA/AL, M3002A, M8102A, M8105A with ECO FW revision E.0 1.22 and 12-lead option C#12 Lots #s included on Consignee list Catalog #: 862442 (M3001A), 866462 (M3001AL), 865039 (M3002A), 865040 (M8102A), 865024 (M8105A)
Why it was recalled
Distorted ST-Segment when using 12-Lead ECG Monitoring with Philips IntelliVue Patient Monitors
Root cause (FDA determination)
Software design
Action the firm took
Philips sent an Urgent Medical Device Correction letter to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that Philips will provide software update and is provding interim instructions for the operator. For questions contact your local Philips representative.
Recalling firm
- Firm
- Philips Electronics North America Corporation
- Address
- 3000 Minuteman Rd, Andover, Massachusetts 01810-1032
Distribution
- Distribution pattern
- Worldwide Distribution and US Nationwide
Timeline
- Recall initiated
- 2017-08-23
- Status
- —