FDA Device Recalls

Recalls /

#160911

Product

Medfusion Syringe Pump 4000 Series The Medfusion Syringe Infusion Pumps are indicated for the following uses: 1) Administration of fluids requiring precisely controlled infusion rates including blood or blood products, lipids, drugs, antibiotics, enteral solutions and other therapeutic fluids; 2) By the following delivery routes: arterial, epidural, intravenous, intrathecal, subcutaneous, and enteral; 3)By the following delivery modes: continuous, volume/time, mass, body weight, intermittent, and bolus; 4) In critical care, anesthesia, neonatal, and pediatric applications or other healthcare settings where use of the syringe infusion pump can be monitored or supervised by a clinician; 5) Inside the MRI room mounted outside the 150 Gauss line and with shielded magnets of field strength of 1.5 Tesla.

FDA product code
FRNPump, Infusion
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K111386
Affected lot / code info
Product Codes: 4000-0101-50, 4000-0101-51, 4000-0105-50, 4000-0105-51, 4000-0150-78, 4000-0106-00, 4000-0106-01

Why it was recalled

Certain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe (i.e. the size of the syringe is unknown to the pump) results in an inability to complete pump programming. Misidentification of a syringe is where the pump misinterprets the syringe size.

Root cause (FDA determination)

Process change control

Action the firm took

Consignees were notified via mail on 11/13/2017, 01/09/2018, and 02/12/2018, as a result of expansions of the recall and the need to provide additional instructions and information to consignees. Instructions in the first letter date 11/13/2017 instructed customers to always confirm the accuracy of infusion values to the original order, including verifying syringe size, to utilize the enclosed Reference Tool when verifying syringe size, and to complete the return the Response Form. Instructions in the letter dated 01/09/2018 included to always confirm the accuracy of infusion values to the original order, including verifying syringe size, to advise all clinicians overseeing use of the affected devices to utilize the enclosed Reference Tool, to remove the pump if the syringe is not recognized or is read as an incorrect size during programming or infusion, to notify impacted customers if the devices were further distributed, and to complete the return the Response Form. Instructions in the letters dated 02/12/2018 included to identify any affected devices on hand, complete and return the Urgent Medical Device Recall Response Form, and arrange for repair of the affected as per the location of the consignee in Canada or the USA. If the consignee is located in the USA, the three options include to return to Smiths Medical for repair, arrange for the shipment of the required parts and have the consignee's onsite Biomed Team complete the repair, or arrange for onsite repair by Smiths Medical if there are more than 10 pumps onsite. If the consignee is located in Canada, the two options include to return to Smiths Medical for repair or arrange for the shipment of the required parts and have the consignee's onsite Biomed Team complete the repair. Consignees not located in the USA or Canada were instructed to contact their Smiths Medical Representative to discuss repair options. For further questions, please send inquiries to SmithsMedical3033@stericycle.com.

Recalling firm

Firm
Smiths Medical ASD Inc.
Address
6000 Nathan Ln N, Minneapolis, Minnesota 55442-1690

Distribution

Distribution pattern
Worldwide distribution- US nationwide Distribution and to the countries of : Australia, Barbados, Canada, Hong Kong, Indonesia, Jamaica, New Zealand, Philippines, Saudi Arabia, Singapore, South Africa, Trinidad and Tobago, United Arab Emirates, and United Kingdom.

Timeline

Recall initiated
2017-11-13
Posted by FDA
2018-03-05
Terminated
2025-09-16
Status

Source: openFDA Device Recall endpoint. Recall record ID #160911. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.